System Specifications; Oracle System Specifications - CEM ORACLE Manual

Rapid nmr fat analyzer
Hide thumbs Also See for ORACLE:
Table of Contents

Advertisement

System Specifications

ORACLE System Specifications

Moisture Range
Fat Range
Balance Capacity
Standard Software
Display
Accessory Ports
Printer (optional)
Dimensions
Weight
Ambient Air Temperature
Voltage
Power
ORACLE Magnet & Processor
Power Conditioner Requirements
RF Pulse Generator
Magnet
Patents
0.01 to 99.99% in all food types, when ORACLE is coupled with SMART 6
0.01% to 99.99% in all food types
100 grams, 0.001g readability
Software with proprietary algorithms
10.1" integrated glass capacitance touch-screen
6 USB, 2 Ethernet
External
ORACLE Magnet
15.6 in (39.6 cm) w x 22 in (56 cm) d x 14 in (35.5 cm) h
ORACLE Processor
12 in (30 cm) w x 13 in (33 cm) d x 16 in (41) h
Processor may be located up to 9'8" (3 m) away from the magnet.
ORACLE Magnet
120 lbs (55 kg)
ORACLE Processor
37 lbs (17 kg)
15 °C and 30 °C - For optimum stability and performance, the ambient
temperature should not vary more than 5 °/day. The ORACLE should not
be near areas with excessive air flow, such as under a fan or air vent.
100-240 V (50-60 Hz)
800 VA maximum. UPS recommended, if power is not reliable.
6.3 amps (100-240 V, 50/60 Hz)
2.1 kVA 120 V, 60 Hz
1.8 kVA 240 V, 50Hz
Pulse Power 300 W nominal. Pulse times variable in 100 ns increments,
Transmit and receive phases selectable 0°, 90°, 180° and 270°. Nominal
90° pulse times 3 ms (10 mm probe) and 4.5 ms (18 mm probe)
Permanent, thermally stabilized, 0.55 T (23.4 MHz proton)
7,581,876 7,997,790 6,227,041 6,787,362 8,530,239 6,548,303
6,548,304 6,268,570 6,320,170 6,521,876 7,125,721 7,220,591