CEM DT-610B Operating Instructions

CEM DT-610B Operating Instructions

Digital thermometer
Hide thumbs Also See for DT-610B:

Advertisement

Quick Links

OPERATING INSTRUCTIONS
DIGITAL THERMOMETER
MAX HOLD
+
BACK-
LIGHT
Tc
C
¡ ã
INTRODUCTION
This instrument is a portable 3
digital thermometer designed to use external K-type
thermocouple
as
temperature
indication follows National Bureau of Standards and
IEC584
temperature/voltage
thermocouples. One K-type thermocouples is supplied
with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and operation
instructions before using the thermometer.
WARNING
TO AVOID ELECTRICAL SHOCK. DO NTO USE THIS
INSTRUMENT
WHEN
MEASUREMENT SURFACE EXCEED 24V AC OR 60V
DC.
WARNING
TO AVOID DAMAGE OR BURNS. DO NOT MAKE
TEMPERATURE MEASUREMENTS IN MICROWAVE
OVENS.
CAUTION
Repeated sharp flexing can break the thermocouple leads.
To prolong lead life, avoid sharp bends in the leads,
especially near the connector.
The
symbol on the instrument indicates that the
operator must refer to "Input protection" in this manual.
¡ ã
C
¡ ã
F
K
0. ¡ ã
1
1 ¡ ã
K
¡ ã
F
1
/
digit., compact-sized
2
sensor.
Temperature
tables
for
K-type
VOLTAGES
AT
THE
SPECIFICATIONS ELECTRICAL
Temperature Scale:
℃ )
Celsius (
, Fahrenheit (
user-selectble
Measurement Range:
-50℃ to 1300℃, -58℉ to 2000℉, 223K to 2000K
Resolution: 1℃ or 1℉, 0.1℃ or 0.1℉,1 K
Accuracy:
Accuracy is specified for operating temperatures over the
range of 18℃to 28℃(64℉ to 82℉), for l year, not including
thermocouple error.
±2℃---------------------- -50℃ to 0℃
±4℉---------------------- -58℉ to 32℉
±5K---------------------- 223K to 273K
±(0.5% rdg+1℃)-------- 0℃ to 1000℃
±(0.8% rdg+1℃)-------- 1000℃ to 1300℃
±(0.5% rdg+2℉)-------- 32℉to 2000℉
±(1.0% rdg+2K)-------- 273Kto 2000K
Temperature Coefficient:
0.1 times the applicable accuracy specification per ℃
from 0℃ to 18℃ and 28℃ to 50℃ (32℉ to 64℉ and
82℉ to 122℉).
Input protection:
60V dc or 24V rms ac maximum input voltage on any
combination of input pins.
Reading Rate: 2.5 times per second.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to center).
ENVIRONMENTAL
Ambient Operating Range:
0℃ to 50℃ (32℉ to 122℉)
Storage Temperature:
-20℃ to 60℃ (-4℉ to 140℉)
Relative Humidity:
0% to 80% (0℃ to 35℃) (32℉ to 95℉)
0% to 70% (35℃ to 50℃) (95℉ to 122℉)
GENERAL
Display:
1
3
/
digit liquid crystal display (LCD) with maximum
2
reading of 1999
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22)
Dimensions:
162mm (H)×76mm(W)×38.5mm(D)
Weight: 210g
Supplied Probe:
4 foot type"K"thermocouple bead probe (Teflon tape
insulated).
Maximum insulation temperature 260℃(500℉)
Probe accuracy ± 2.2 ℃
(Whichever is greater) from 0°to 800℃
OPERATING INSTRUCTIONS
Selecting the Temperature Scale
Readings are displayed in either degrees Celsius (℃),
degrees Fahrenheit (℉) or degrees
thermometer is turned on, it is set to the temperature scale
that was in use when the thermometer was last turned off.
To change the temperature scale, press the For℃,℉or K
key.
℉ )
or Kelvin (
K)
or ± 0.75% of reading
Kelvin
(K).When the

Advertisement

loading

Summary of Contents for CEM DT-610B

  • Page 1 OPERATING INSTRUCTIONS SPECIFICATIONS ELECTRICAL Temperature Scale: ℃ ) ℉ ) Celsius ( , Fahrenheit ( or Kelvin ( DIGITAL THERMOMETER user-selectble Measurement Range: -50℃ to 1300℃, -58℉ to 2000℉, 223K to 2000K Resolution: 1℃ or 1℉, 0.1℃ or 0.1℉,1 K Accuracy: MAX HOLD Accuracy is specified for operating temperatures over the...