Advertisement

5. Insert the CEM USB located in the Normalization Kit into the ORACLE Processor Box USB Port.
NOTE
Ensure the NS number on the Normalization tube matches that on the USB flash drive. If the two numbers
do not match, please contact CEM Corporation.
6. Once the USB is plugged into the ORACLE Processor Box, highlight the USB drive listed in the window and
select Import to complete the process.
7. Once the Clone file has been imported, select "Auto Run" to perform a full System Setup.
8. The ORACLE will prompt the user to "Please insert Oil standard tube" in the magnet. Place the glass Oil Stan-
dard tube into the magnet. Ensure the Oil Standard tube is completely placed down in the ORACLE by gently
pressing it down on the top.
9. Select "continue".
Software Setup
27