Signal Identification - Agilent Technologies 11970 Series User Manual

Harmonic mixers
Hide thumbs Also See for 11970 Series:
Table of Contents

Advertisement

Operation
Using the Mixers with the 8566B Spectrum Analyzer

Signal Identification

Figure 2-18 on page 55
shows a typical full-band display of a single input signal. Direct interpretation of
the display is difficult because of the large number of responses produced by several local oscillator
harmonics generated in the mixer. To solve this problem the 8566B spectrum analyzer uses two methods
of signal identification.
The first is an automatic image search in which the spectrum analyzer looks for a signal image within a
preset range of harmonic numbers. This automatic signal identification routine is initiated by pressing
,
,
. Upon completion, one of three messages appears on the CRT: SIGNAL NOT
SHIFT
3
MHz
IDENTIFIED, SIGNAL IDENTIFIED OUT OF BAND (with the frequency), or CENTER (with the
identified signal frequency). The range of harmonic numbers through which the spectrum analyzer
searches can be changed by accessing "search start HN" and "search stop HN," which are
,
,
SHIFT
7
MHz
and
SHIFT
,
8
,
MHz
, respectively.
The second method, manual signal identification, is a fast method of verifying that the displayed signal is
a true signal in the specified band. Press
SHIFT
,
4
,
MHz
to activate this function. Positive signal
identification is indicated if the signal in question does not change frequency and is reduced in amplitude
by one to two divisions on alternate sweeps. A failed signal identification is indicated by the signal
disappearing or shifting in frequency.
56

Advertisement

Table of Contents
loading

This manual is also suitable for:

11970k11970a11970q11970u11970v11970w

Table of Contents