Summary of Contents for Agilent Technologies 1141A
Page 1
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 3
Each of the two instrument models that make up the differential probe system has a serial number sticker. The sticker for the 1141A Differential Probe is inside the probe, in the bottom cover. (See chapter 3 for disassembly procedure.)
Contents 1 Operating the Probe Accessories Supplied 6 Accessories Available 8 To inspect the probe 9 Using the probe with other instruments 9 Recommended Test Equipment 9 2 Calibration Tests and Adjustment Equipment Required 26 The Test Board 26 Calibration Tests 27 dc Gain Accuracy 28 Bandwidth 34 CMRR Test 36...
Page 5
Contents Theory of Operation 59 Differential Probe 60 Control and Power Module 60 Attenuator Adapters 61 Test Board 62 Service Policy 63 Troubleshooting 64 Probe Troubleshooting 64 Probe Control and Power Module Troubleshooting 64 Removing and Replacing Assemblies 66 Differential Probe 66 Probe Adapters 68 Probe Control and Power Module 70 Replaceable Parts 71...
Operating the Probe Accessories Supplied Introduction This chapter shows you how to connect and operate the 1141A Differential Probe and 1142A Probe Control and Power Module as a differential probe system. Accessories Supplied The following items are supplied as part of the 1141A/1142A probe system.
Page 8
Operating the Probe Accessories Supplied Figure 1-1 1141A Differential Probe and Accessories Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Operating the Probe Accessories Available Figure 1-2 1141A Miscellaneous Accessories Figure 1-3 1142A Probe Control and Power Module Accessories Available The following accessories can be ordered. • 5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5 • 5090-4833 Mini grabber for SMT, package of 20...
If the shipping container is damaged, or the cushioning materials show sign of stress, notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping materials for the carrier’s inspection. The office will arrange for repair or replacement at Agilent Technologies’...
Page 11
The following paragraphs cover system preparation and initial adjustments. Power Requirements The 1141A/1142A probe system (specifically the 1142A) requires a power source of either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum. C A U T I O N Before connecting power to this instrument, be sure the line voltage switch on the rear panel of the instrument is set properly.
Page 12
Initial Adjustment For a given combination of 1141A Differential Probe and 1142A Probe Control and Power Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need to make these adjustments only once, before the probe is first used. You can make them any time to optimize the system.
Page 13
Operating the Probe Recommended Test Equipment • Offset null zeroes the dc level at the output of the probe. The range of adjustment is about ±4 mV. • DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc component of an input signal.
Page 14
Using the Accessories The 1141A Differential Probe and accessories are designed to provide a variety of ways to connect to circuitry and make measurements. In the descriptions, any method used to connect to the probe signal inputs also applied to the adapters. The figure below shows, in a general way, the use of accessories.
Page 15
Operating the Probe Recommended Test Equipment Figure 1-7 Basic Accessory Connections Probe Tips Probe tips fit into the receptacles in the probe and are held in place with probe tip caps. If necessary, you can solder the probe tips into a circuit or wires can be soldered to the tips.
Page 16
To prevent pickup of stray fields when you use extension lead, either the ones supplied with the 1141A or others, dress them carefully as follows: • Connect the leads at right angles to the circuitry under test.
Page 17
Operating the Probe Recommended Test Equipment Adapters There are three adapters for use with the differential probe. Two adapters are attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe input. The adapters are installed on the probe after the probe tip caps and probe tips have been removed.
Page 18
Operating the Probe Recommended Test Equipment Adapter Combinations The figure below shows the allowed adapter and probe connections. There are two specific combinations that should not be used. • Do not attach the ac adapter between an attenuator adapter and the probe. An attenuator adapter must be terminated by the input resistance of the probe.
Page 19
Operating the Probe Recommended Test Equipment Connector Compatibility The following are general connector characteristics for the probe, adapters, and accessories. • The female connectors on the probe, adapters, and other accessories are designed to mate with 0.030-inch round or 0.0250-inch square pins. •...
Page 20
Operating the Probe Recommended Test Equipment Probe System Coupling Functions dc offset dc Reject ac Coupling Adapter dc Blocked ± 20 V ± 20 V ± 20 V Probe alone ± 200 V ± 200 V ± 500 V Probe with 10x ±...
Page 21
Operating the Probe Recommended Test Equipment To use dc reject: Remove the ac adapter if it is installed. On the front panel of the 1142A, press Local. Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or 5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz.
Page 22
Operating the Probe Recommended Test Equipment To use ac coupling: Attach the ac coupling adapter to the input of the probe or the input of the attenuator adapter. On the 1142A, press Local and Zero offset. Remote operation For automatic test applications, the coupling and offset functions provided by the 1142A Probe Control and Power Module can be remotely controlled through a connector on the rear panel of the module.
Page 23
Operating the Probe Recommended Test Equipment Function Select The easiest way to control the function select lines is contact closures between the lines and Digital common (pin 2) of the remote input connector. (TTL compatible control signals can be used; but to avoid problems with ground loops, they must be electrically isolated.) The following truth table shows the functions provided by the function select lines.
Page 24
The CMRR of the 1141A Differential Probe is specified at the input of the probe and cannot be affected expect by adjustments in the probe. However, the way the probe is connected into the circuitry being tested can have a big influence in the overall result of the measurement, especially at high frequencies.
Page 25
Operating the Probe Recommended Test Equipment Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Equipment Required Introduction This chapter is divided into two sections. The first section gives calibration tests and the second adjustment procedures for the 1141A Differential Probe and 1142A Probe Control and Power Module. Equipment Required A complete list of equipment required for the calibration tests and adjustments is listed in “Recommended Test Equipment”...
Calibration Tests and Adjustment The Test Board Calibration Tests These procedures test the probe’s electrical performance using applicable specifications given in “Performance Specifications and Characteristics” on page 55 as performance standards. Specifications applicable to individual tests are noted at the test for reference. Testing Interval The calibration testing procedures may be performed for incoming inspection of the instrument and should be performed periodically thereafter to ensure and maintain peak...
Calibration Tests and Adjustment dc Gain Accuracy dc Gain Accuracy This test checks the dc gain accuracy of the differential probe and the dc accuracy of the differential probe with attenuator and adapters. Specification: Probe alone, ±2%; with attenuator adapter, ±4% Equipment Required Recommended Equipment...
Page 30
Set the Local/Remote push button to Local. Under DC Couple, press the Zero offset button. With the 1141A Probe Amp disconnected from the test PCA, adjust the Offset Null control on the 1142A until the DVM reads 0Vdc. If the probe output voltage cannot be set to 0V, subtract this voltage from the subsequent measurements in this test.
Page 31
Calibration Tests and Adjustment dc Gain Accuracy Figure 2-3 Signal to - input Calculate probe gain as ∆V – ------------- - --------------------------------- - ∆V × Record the result of this calculation in the “Calibration Test Record” on page 39. To pass this test, the probe gain = 0.98 to 1.02 N O T E Failure of the gain accuracy test can be caused by mis-adjustment of the probe.
Page 32
Calibration Tests and Adjustment dc Gain Accuracy Figure 2-4 Signal to + input Set the dc calibrator output 3 V dc. Record the V measurement from the top DVM in figure 2-2. Record the V measurement from the bottom DVM in figure 2-2 out1 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input).
Page 33
Calibration Tests and Adjustment dc Gain Accuracy 100x Attenuator Accuracy Test N O T E If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator adapter. Do not continue until the probe passes the gain test. Disconnect the probe/attenuator from the test board.
Page 34
Calibration Tests and Adjustment dc Gain Accuracy Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input). Figure 2-7 Signal to - input Record the V measurement from the bottom DVM in figure 2-2. out2 Calculate the 100x attenuator gain as ∆V...
Calibration Tests and Adjustment Bandwidth Bandwidth This test checks the high-frequency response of the 1141A Differential Probe. The bandwidth of the oscilloscope is characterized first so it is not a factor in the measurement. Specification (-3dB, dc coupled): dc to 200 MHz...
Page 36
Calibration Tests and Adjustment Bandwidth Reconfigure the equipment. Disconnect the signal generator cable from the oscilloscope input and connect it to the test board. Connect the output of the differential probe to the channel 1 input of the oscilloscope. Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Calibration Tests and Adjustment CMRR Test CMRR Test This test checks the CMRR at 1 MHz and 100 MHz. Specification 3000:1 at 1 MHz, 10:1 at 100 MHz Equipment Required Recommended Equipment Agilent Required Critical Specifications Model/Part Oscilloscope 400 MHz bandwidth at 1 mV/div 54830B Signal Generator 1-100 MHz at = 400 mVrms...
Page 38
Calibration Tests and Adjustment CMRR Test On the oscilloscope, press AUTOSCALE and set the following parameters. Menu Selection Setting TIMEBASE (time/div) 500 ns/div CHAN 1 (sensitivity) 200 mV/div 50 Ω DC (input R) ACQUISITION Sampling Mode Real Time Memory Depth Automatic Sample Rate Automatic...
Page 39
Calibration Tests and Adjustment CMRR Test Connect the input of the probe to the test board in the position shown in the figure below (signal to + input). Figure 2-11 Signal to + input Set the signal generator for 100 MHz at 0.0 dBm (about 224 mV , 632 mV Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div.
Calibration Tests and Adjustment Calibration Test Record Calibration Test Record 1141A/1142A Differential Probe Tested by_________________________ Serial No. ______________________________ Work Order No.____________________ Recommended Test Interval - 1 Year/2000 hours Date____________________ Recommended next testing_________________ Temperature_____________ Test Limits Results dc Gain Accuracy Probe +0.98 mV to +1.02 mV...
Calibration Tests and Adjustment Probe Adjustment Adjustments This section provides adjustment procedures for the 1141A Differential Probe and attenuator adapters. There are no service adjustments for the 1142A Probe Control Module. Adjustment Interval None of the adjustment procedures that follow should be considered for a routine maintenance plan.
Page 42
Calibration Tests and Adjustment Probe Adjustment The following equipment is required for this procedure. Procedures are based on the model or part number recommended. Equipment Required Equipment Required Critical Specifications Recommended Agilent Model/Part Function Generator 2.5 kHz, 1 V 33120A Oscilloscope 300 MHz at 1 mV/div 54830B...
Calibration Tests and Adjustment Adjustment Procedure As shown in the figure below, use the grounding screw to reinstall the ground block on the PC assembly. Figure 2-13 Attaching Ground Block to Probe PC Assembly The ground block provides a mechanical and electrical connection when the probe PC assembly is connected to the test board.
Page 44
Calibration Tests and Adjustment Adjustment Procedure On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set up the following parameters. Menu Selection Setting TIMEBASE (time/div) 500 ns/div CHAN 1 (sensitivity) 100 mV/div 50 Ω DC (input R) TRIG (mode) trg’d source level...
Page 45
Calibration Tests and Adjustment Adjustment Procedure Center adjustment R11, HF CMRR (see following figure). Figure 2-15 R11, HF CMRR Adjustment Adjust R9, HF GAIN so the V (1) measurement is the same as in step 4, within ±0.5%. Make the adjustment slowly so the oscilloscope display has time to react to signal averaging.
Page 46
Calibration Tests and Adjustment Adjustment Procedure Change the oscilloscope settings to: Menu Selection Setting TIMEBASE (time/div) 50 µs/div CHAN 1 (sensitivity) 100 mV/div Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input). Figure 2-17 Signal to + input Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top...
Page 47
Calibration Tests and Adjustment Adjustment Procedure Carefully connect the probe to the test board in position shown in the figure below (signal to both inputs). Figure 2-19 Signal to both inputs Change the function generator settings to: • Sine wave •...
Page 48
Calibration Tests and Adjustment Adjustment Procedure High Frequency Compensation This adjustment sequence continues from the Low Frequency Response and CMRR adjustment. However, it can be done separately if the probe meets all specifications except bandwidth. Adjust R13 for unity gain at 200 MHz. Connect the signal generator to the test board and set it for 200 MHz and 300mV (107 mV...
Page 49
Calibration Tests and Adjustment Adjustment Procedure Assemble the PC assembly in the top cover. The side of the PC assembly with the large hybrid is exposed when the assembly is in the top cover. The figure below shows how the top cover, PC board, and ground block fit together.
Attenuator adapters have only characteristics; they do not have any specifications. An adapter will need adjustment only if one or more of the following occurs. • If an adapter is to be used on a different 1141A probe that it was calibrated with last.
Page 51
Adjustment Procedure N O T E The attenuator must be adjusted when installed on the 1141A probe with which it will be used. The specifications and characteristics will not be met if the attenuator adapter is adjusted with one differential probe then used with another.
Page 52
Calibration Tests and Adjustment Attenuator Adapter Adjustment Connect the adapter/probe combination to the test board in the position shown in the figure below. Figure 2-23 Signal to + input Adjust the + HF RESP for best overall pulse response, the flattest pulse top. Use the figure below for adjust locations.
Page 53
Calibration Tests and Adjustment Attenuator Adapter Adjustment Connect the adapter/probe combination to the test board in the position shown in the figure below (signal to both inputs). Figure 2-25 Signal to both input Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal on the oscilloscope.
Introduction This section provides troubleshooting, service, and repair information for the 1141A Differential Probe and 1142A Probe Control and Power Module. The troubleshooting information is provided to isolate a faulty assembly. When a faulty assembly has been located, the disassembly/assembly procedures help direct replacement of the assembly.
Service Performance Specifications and Characteristics Performance Specifications and Characteristics The following table gives performance specifications used to test the 1141A and 1142A. It also gives performance characteristics that are typical for the probe system. Performance Specifications and Characteristics Parameter Probe alone...
Page 57
Service Performance Specifications and Characteristics Figure 3-1 CMRR Specifications and Characteristics Legend CMRR specification for probe with no input adapters. Low-frequency CMRR specification for probe with the ac adapter. Typical CMRR characteristic for differential probe with no input adapters Typical CMRR characteristic for differential probe with 100x attenuator adapter at input. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
Page 58
Service Performance Specifications and Characteristics Figure 3-2 Maximum Input Voltage vs. Frequency Legend Input voltage limits for probe alone. Input voltage limits for 10x adapter. Input voltage limits for 100x adapter. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Service General Characteristics General Characteristics The following characteristics apply to the 1141A Differential Probe with the 1142A Probe Control and Power Module. Environmental Conditions Operating Non-operating Temperature 0°C to +55 C° (32°F to +131°F) -40°C to +70°C (-40°F to +158°F)
The following discussion covers block-level theory for the 1141A/1142A differential probe system. Refer to the block diagram below. The differential probe system consists of two units, the 1141A Differential Probe with its accessories and the 1142A Probe Control and Power Module. For purposes of the following discussion, these will be called the probe and the control module respectively.
Service Theory of Operation Differential Probe The probe contains a two-path differential amplifier with unity gain. It is implemented on a double-sided surface-mount PC board with the high-frequency path on one side and the low-frequency path on the other. The two paths are split directly after the differential input connections.
Service Theory of Operation Offset Functions There are two offset functions developed in the control module: variable offset and dc reject. A variable offset voltage with coarse and fine adjustments can be selected by the front panel controls. The offset level is buffered by U8 and selected by multiplexer U3 as the input to offset amp U7.
Service Theory of Operation Test Board The test board is a device for conveniently connecting test signals to the differential probe. The probe can be connected to the board with the signal to the positive, negative, or both inputs. Figure 3-6 Test Board Schematic Artisan Technology Group - Quality Instrumentation ...
Service Service Policy Service Policy For parts of the 1141A/1142A probe system that are complex, the service policy is for assembly-level repair. For parts of the system with simple circuitry, the service policy is component-level repair. The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies include the PC assembly and cable.
Service Troubleshooting Troubleshooting Use the following paragraphs to assist in troubleshooting problems with the 1141A/1142A Differential Probe. Probe Troubleshooting To troubleshoot the probe: Apply a known signal to the input of the probe. Check for an identical output at the output coax to the cable. This connection is the one soldered to the PC board.
Page 66
Service Troubleshooting Figure 3-7 1142A Probe Control and Power Module Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Service Removing and Replacing Assemblies Removing and Replacing Assemblies This section contains procedures for the removal and replacement of major assemblies. C A U T I O N Never remove or install any assembly with the instrument power ON. Component damage can occur.
Page 68
Service Removing and Replacing Assemblies Reassemble Probe The ground screw passes through the top cover and PC board and screws into the ground block. If replacing the PC board, remove the input connectors from the old board and put them on the new one. If replacing the cable, note the orientation of the probe clamp ring on the old cable, remove the ring and put it on the new cable.
Service Removing and Replacing Assemblies C A U T I O N Note where the two pins at the rear of the top cover enter the holes in the PC assembly. Position the cable wires away from these two areas. When the bottom cover is closed, part of it will pinch wires that are laying over these areas.
Page 70
Service Removing and Replacing Assemblies Combine the thumbwheel screw and thumbwheel and insert them into the hole in the outer housing. Insert the substrate/board into the outer housing. Slip the attenuator ground (attenuator adapters only) over the thumbwheel screw and seat the input connectors in the proper holes in the housing.
Service Removing and Replacing Assemblies Probe Control and Power Module Use the following procedure to disassemble the probe control and power module. W A R N I N G Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely to the following procedures.
• Prepaid transportation (there is a small handling charge for each order). • No invoices. In order for Agilent Technologies to provide these advantages, a check or money order must accompany each order. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Service Replaceable Parts Mail order forms and specific ordering information are available through your local Agilent Technologies Sales Office. Addresses and telephone numbers are located in a separate document shipped with the manuals. Manufacturers’ Codes A list of manufacturers’ codes is given the table below. The codes are given for parts in the parts lists.
Service Replaceable Parts Exploded View Figure 3-11 1141A Differential Probe Parts Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 75
Service Replaceable Parts Figure 3-12 Power and Control Board Component Locator Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 76
Service Replaceable Parts Figure 3-13 Cabling Diagram Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 77
Service Replaceable Parts Figure 3-14 IC Connectors not Shown Supply Pin No. IC Group U6-8, 11 1, 5, 8 1142A Probe Control and Power Module Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 78
Service Replaceable Parts Figure 3-15 1142A Probe Control and Power Module Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 79
Service Replaceable Parts 1141A and 1142A Replaceable Parts Ref. Part Description Mfr. Mfr. Part Des. Number Code Number 1141A Differential Probe 01141-66503 PC ASSEMBLY-DIFFERENTIAL PROBE 28480 01141-66501 01141-26102 PROBING PIN 28480 01141-26102 01141-22301 GROUND BLACK 28480 01141-22301 01141-22401 GROUND SCREW...
Page 80
Service Replaceable Parts 1141A and 1142A Replaceable Parts Ref. Part Description Mfr. Mfr. Part Des. Number Code Number 01142-66501 PC ASSEMBLY-POWER AND CONTROL 28480 01142-66501 0515-0374 SCREW-MACHINE M3 10mm-LG 00000 ORDER BY DESP. 0515-1031 SCREW-MACHINE M3 6mm-LG 90-DEG-FLH-HD 00000 ORDER BY DESP.
Page 81
Service Replaceable Parts Power Supply Replaceable Parts Ref. Des. Part Description Mfr. Mfr. Part Number Code Number Prefix the reference designators with A1 0160-6190 CAPACITOR-FXD 0.33UF ±10% 50VDC 28480 0160-6190 0160-5474 CAPACITOR-FXD 0.1UF ±5% 100VDC MET-POLY 28480 0160-5474 0180-3298 CAPACITOR-FXD 2200UF+30-10% 50VDC AL 28480 0180-3298 0180-3298 CAPACITOR-FXD 2200UF+30-10% 50VDC AL...
Page 82
Service Replaceable Parts Power Supply Replaceable Parts Ref. Des. Part Description Mfr. Mfr. Part Number Code Number NOT ASSIGNED 1901-1087 DIODE-PWR RECT 100V 3A 04713 MR501 1901-1087 DIODE-PWR RECT 100V 3A 04713 MR501 NOT ASSIGNED 1901-0731 DIODE-PWR RECT 400V 1A 28480 1901-0731 1901-0731 DIODE-PWR RECT 400V 1A...
Page 83
Service Replaceable Parts Power Supply Replaceable Parts Ref. Des. Part Description Mfr. Mfr. Part Number Code Number 1252-4731 CONNECTOR-AC PWR 28480 1251-4743 1252-1487 CONN-RECT D-SUBMIN 9-CKT (remote) 28480 1252-1487 1252-3935 CONNECTOR-ROUND 8-CKT (probe) 28480 1252-3134 1400-1604 LED MOUNT 28480 1400-1604 1205-0732 SPRING CLIP 28480 1205-0732...
Page 84
Service Replaceable Parts Power Supply Replaceable Parts Ref. Des. Part Description Mfr. Mfr. Part Number Code Number 0698-6317 RESISTOR 500 0.1% 0.125W TF TC=0±25 28480 0698-6317 0698-4002 RESISTOR 5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5001-F 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203 0698-6317 RESISTOR 500 0.1% 0.125W TF TC=0±25...
Page 85
Service Replaceable Parts Power Supply Replaceable Parts Ref. Des. Part Description Mfr. Mfr. Part Number Code Number 1810-1242 RESISTOR NETWORK 28480 1810-1242 3101-2609 SWITCH-SL DPST STD 5A 250VAC PC 28480 3101-2609 3101-3007 SWITCH-6 STATION ASSEMBLY 28480 3101-3007 9100-4750 TRANSFORMER-POWER (with mtg. hardware) 28480 9100-4750 1826-1403 IC V RGLTER-ADJ-POS 3/40V...
Page 86
Service Replaceable Parts Locator Table for Control and Power Supply Ref. Grid Grid Grid Grid Grid Grid Grid Grid Des. C-6 CR10 C-5 B-5 CR11 C-3 B-6 CR12 C-6 B-5 CR13 C-3 B-5 CR14 C-6 B-6 CR17 B-5 B-4 CR18 B-6 B-4 DS1 C-3 F1 C-3 HS1...
Page 87
Service Replaceable Parts Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 88
Index Numerics 100x attenuator test cabling diagram environmental characteristics 10x attenuator test calibration exchange assemblies test record extension leads tests characteristics ac adapter environmental ac coupling adapters features general ac low frequency frequency performance accessories high compensation probe connectors available to order low response circuit connector posts circuit connector posts...
Page 89
Index probe gain test procedures, calibration probe preparation test record, calibration making measurements testing interval manufacturer’s codes theory of operation maximum input voltage thermal drift mini grabbers recommended test equipment troubleshooting record calibration power module remote control probe remote functions noises probe control remote input connection...
Page 90
7015 (b)(2) (November 1995), as information contained herein, agreement and written consent applicable in any technical data. including but not limited to the from Agilent Technologies, Inc. as implied warranties of governed by United States and Manual Safety Notices merchantability and fitness for a international copyright laws.
Page 91
Product Safety any unintended operation. Safety Symbols Notices Service instructions are for trained service personnel. To avoid dangerous electric shock, This apparatus has been designed Instruction manual symbol: the do not perform any service unless and tested in accordance with product is marked with this qualified to do so.
Page 92
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...