In this example, we'll look at U-band, which ranges from 40 GHz to 60 GHz, as shown in
The
LOCK HARMONIC
analyzer sweeps only the chosen band.
pressed.
Figure 2-9
Select the band of interest.
Save the average conversion-loss value
4. Default conversion-loss values that are stored in the analyzer for each frequency band are listed in
Table 2-1 on page
40. These values approximate the values for the Agilent 11970 series mixers. Other
conversion-loss values may be entered into the spectrum analyzer in two ways. The first method lets
you to save the average conversion-loss value for the entire band using
activate this function:
a. Press
AUX CTRL
b. Enter the appropriate average conversion-loss value. On Agilent 11970 Series harmonic mixers,
these values are listed on the mixer, or a calibration sheet that is supplied with the mixer. The
11970A, Q, and V band mixers calibrated after May 2011 have two calibration sheets. Use the
sheet that contains the text "Not for use with the Agilent X-Series signal analyzers". The U-band
mixer used for this example had an average conversion loss of 23.5 dB, as shown in
on page
42.
Using the Mixers with the 856X Series Spectrum Analyzers
function "locks" the spectrum analyzer in that band, ensuring that the spectrum
LOCK ON OFF
,
,
EXTERNAL MIXER
AMPTD CORRECT
is automatically set to ON when
AVERAGE CNV LOSS
,
AVERAGE CNV LOSS
Operation
Figure
2-9.
FULL BAND
is
. To
.
Figure 2-10
41