JEOL JNM-ECA Series User Manual page 53

Solid measurement
Table of Contents

Advertisement

3.
Click the Set Value button.
The array parameter window closes, and the values you set are entered into
tau_interval in the Experiment Tool.
4.
Click the Submit button.
The measurement starts.
Although the data-processing procedure is the same as that for the inversion-recovery
method, use the following approximation formulas.
Weighted Linear Sat. Recovery:
Unweighted Linear Sat. Recovery: Linear least-squares method without weights.
Nonlinear Saturation Recovery:
Refer to the user's manual "Application" for information about the measurement and
processing of the inversion-recovery method.
NMECAXS_V50-SLD-2
6 MEASUREMENT OF RELAXATION TIME
Fig. 6.1 Set interval window
The weighted linear least-squares method
The weight is reduced for the measurement having
long τ.
Nonlinear least-squares method
Arrow buttons
6-3