CALIBRATION BY USB
The F6.61 hot tap magmeter can be connected to a PC and the operator can
be connected to a PC and the operator can calibrate the instrument and set
all parameters using dedicated software on USB pen drive (software can be
dowloaded freely from FLS website)

Procedure for setting

- plug FLS Pen Drive into a USB port of PC
- open FLS Pen Drive folder
- install Java software: a internet connection is required. Installer software is
able to search for updated Java version automatically. In case of troubles,
please contact Technical Assistance at info@flsnet.it
- install FLS Calibration software
- launch FLS Calibration software
- set all MicroSwitches in ON position
- power the F6.61
- plug USB on the F6.61 PCB
- plug USB into a USB port of PC
- FLS Calibration software recognizes the F6.61 PCB

Software structure

The FLS Calibration software features following sub-views:
• Settings
• Calibration
• Output mA
• Digital Output
• Simulation
• View Data
• Download Data
Excluding View Data Section and Download Data Section, in each previous
sub-view you can set different parameters and the following actions are
allowed:
- Update: for data updating
- Reset: for going back to default data
- Help: for functions explanation and for undertaking the Remote Assistance
procedure (internet connection is required)
View Data sub-view summarizes status of measurement, analog output, digital
output including Volume Totalizer .
Download Data sub-view summarizes instrument parameters and following
actions are allowed:
- Download Data: to update instrument with new settings
- Download Default: to reload Default settings
- Save: to generate a file containing all set parameters
- Load: to load directly a file containing a instrument configuration