Disinfecting

If you plan to use other disinfectants, approval of Agfa is needed before use, as
most disinfectants can damage the device. UV disinfection is also not allowed.
Perform the procedure following the instructions for use, the disposal
instructions and the safety instructions of the selected disinfectants and tools
and of the hospital.
Items contaminated with blood or body fluids, which may contain blood-
borne pathogens, should be cleaned and then receive intermediate level
disinfection with a product having an EPA-registered claim for activity against
hepatitis B.
DR 10e, DR 14e, DR 17e | Introduction to the DR Detector | 51
WARNING:
To disinfect the device, use only disinfectants and disinfection
methods that are approved by Agfa and that correspond to the
national regulation and guidelines as well as explosion
protection.