38 | Drystar 5301 XL, Drystar 5302 XL | Advanced Operation (key-operator mode)
5. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
Related Links
Accessing the Change settings menu
Understanding the calibration policy
Changing general image quality settings - Default densitometer
1. On the Change settings menu, press the Down key four times, followed by
the Confirm key to select 'Image quality'.
The following screen will appear:
2 <Modality name>
3 <Modality name>
2. Press the Confirm key to select 'General settings'.
The 'General setting' screen is displayed:
3. Press the Down key once, followed by the Confirm key to select 'Def.
densitometer'.
The 'Default densitometer' screen is displayed:
1 Macbeth TR924
2 Gretag
3 X-Rite301
4*X-Rite310
5 X-Rite331
6 X-Rite341
The current default densitometer is marked with an *.
4. Press the Up/Down arrow keys to select the default densitometer.
5. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
3730D EN 20210520 0952
Note:
1 pack = 100 sheets
E.g., if the frequency is set to 5, this means that a film
calibration is executed every 5 packs or every 500 sheets.
Note: You can specify to perform an automatic film
calibration after loading every "x" film packs.
on page 23
on page 127