Positioning the Bucky

The bucky center position is automatically aligned to the position of the X-ray
tube stand. The mechanical coupling between the bucky and the X-ray tube
stand is active within the travel range of the bucky.
The bucky can be positioned independently from the X-ray tube stand, e.g. for
oblique X-ray exposures.
To position the bucky independently from the X-ray tube stand:
1. Move the X-ray tube stand on the longitudinal axis outside the travel range
of the bucky.
The mechanical coupling is released.
2. Press and hold the bucky lock switch.
Figure 33: Bucky lock switch
The lock for the bucky movement is released.
3. Move the bucky in longitudinal direction.
4. Release the bucky lock switch.
The position is locked.