COMMISSIONING

Connect the drive unit to the electrical supply. Check the supply voltage against the rating plate.
Immerse the dispersing unit in the product. The optimum depth of immersion of the dispersion unit is about
2/3 of the total depth of product in the container. Never immerse the upper holes of the dispersion attachment.
Set the desired speed.
Note:
In some cases, it is better to position the dispersion unit off centre in order to obtain more turbulence and better
mixing. This can be achieved by tilting the drive by about 15°.
Attention
13
vwr.com I Instruction manual VWR
Never allow the VWR Homogeniser dispersion attachment to run without a liquid
phase. Running dry will destroy the sleeve bearing.
Ensure that the dispersion attachment is thoroughly cleaned after every use.
Never touch the rotor of a VWR Homogeniser dispersion attachment when rotating.
There is great danger of injury.