Electromagnetic Environmental Recommendations; (Disposable Sets); Disposable Set Accessories - Smiths Medical Level 1 NORMOFLO H-1129 Operator's Manual

Irrigation fluid warmer
Table of Contents

Advertisement

Electromagnetic Environmental Recommendations

Recommended separation distances between portable and mobile RF communications equipment and
the Level 1
®
The Level 1
NORMOFLO
which radiated RF disturbances are controlled. The customer or the user of the Fluid Warmer can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Fluid Warmer as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 1 The guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Level 1
®
NORMOFLO

(Disposable Sets)

Product Description
IR-40
Normothermic Irrigating Set
IR-500
Normothermic Irrigating Set
IR-600
Normothermic Irrigating Set
IR-700
Normothermic Irrigating Set with vented spikes for use with
rigid plastic containers

Disposable Set Accessories

Product Description
IR-6
2-spike adapter to convert IR/IRI series disposables to a
4-spike configuration
IR-7
2-spike adapter to convert IR/IRI series disposables to a
4-spike configuration - for use with rigid plastic containers
Level 1
®
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer is intended for use in an electromagnetic environment in
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d=[3.5/V1]√P
0.116
0.368
1.16
3.69
11.66
®
Irrigation Warming Sets
®
®
NORMOFLO
H-1129 Irrigation Fluid Warmer | Operator's Manual
S E C T I O N 1 5
H-1129 Irrigation Fluid Warmer
80 MHz to 800 MHz
d=[3.5/E1]√P
0.116
0.368
1.16
3.69
11.66
S p e c i f i c a t i o n s
800 MHz to 2.5 GHz
d=[7/E1]√P
0.233
0.737
2.33
7.38
23.33
51