A–Install the Irrigation Warming Set
The installation sequence for the Disposable Set corresponds to the
numbered Blocks marked 1-2-3 on the Fluid Warmer.
Remove the Disposable Set from its packaging and review the
Instructions for Use provided. Do not remove spike caps at this time.
1 Push the bottom end of the Heat Exchanger ( a ) [the end near the
Gas Vent] into #1 Block ( b ). Press the Heat Exchanger down
firmly to properly seat in the block.
2 Slide #2 Block up ( c ). Snap Heat Exchanger into guide ( d ). Press
firmly into place to ensure it is properly seated. Slide #2 Block
down ( e ), push down firmly to secure.
®
®
Level 1
NORMOFLO
H-1129 Irrigation Fluid Warmer | Operator's Manual
S E C T I O N 9
•
O p e r a t i n g I n s t r u c t i o n s
a
a
b
c
e
d
29