Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Guidelines and manufacturer's declaration – electromagnetic immunity
The LED battery light source, models 11301 D1, 11301 D3 and 11301 D4, is intended for use in an electromagnetic environment specified below. The user of
the LED battery light source, models 11301 D1, 11301 D3 and 11301 D4, should make sure that it is used in such an environment.
Immunity tests
Electrostatic discharge (ESD) acc.
± 8 kV contact discharge
to IEC 61000-4-2
± 15 kV air discharge
Rapid transient electrical
± 2 kV for power lines lines
Interferences/bursts acc.
± 1 kV for input and output lines
to IEC 61000-4-4
100 kH repetition
Surges
± 1 kV line(s) to line(s)
acc. to IEC 61000-4-5
± 2 kV line(s) to earth
Voltage Dip:
Dip to 0% for 1 cycles @
0°phase angle
Dip to 70% for 25/30 cycles @
0° phase angle
Voltage dips, short interruptions
and fluctuations of the supply
Dropout to 0% for 0.5 cycles @
voltage acc. to IEC 61000-4-11
0°, 45°, 90°, 135°, 180°, 225°,
270° & 315° phase angles
Voltage interruption:
100% for 250/300 cycles
Magnetic field at the
supply frequency (50/60 Hz)
30 A/m at 50Hz/60Hz
acc. to IEC 61000-4-8
3V/m 80 MHz to 2.7 GHz
Radiated, radio-frequency,
electromagnetic field immunity test
*Refer table 2 for wireless
acc. to IEC 61000-4-3
Proximity RF field test levels
Immunity to conducted
3Vrms on 150 kHz to 80MHz
disturbances, induced by radio-
1kHz 80% AM modulation
frequency fields
6Vrms in ISM bands
acc. to IEC 61000-4-6
a
These tests are not applicable as the device does not have a connecting cable.
Information on
electromagnetic
compatibility (EMC)
Table 1
EN/IEC 60601 test level
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
not applicable
not applicable
not applicable
30 A/m at 50Hz/60Hz
3V/m 80 MHz to 2.7 GH
not applicable
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
a
commercial or hospital environment.
Mains power quality should be that of a typical
a
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the LED battery light source, models
a
11301 D1, 11301 D3 and 11301 D4, requires
continued operation during power supply interruptions,
it is recommended that the device be powered from an
uninterruptible power supply or a battery.
If image distortion occurs, it may be necessary
to position the LED battery light source, models
11301 D1, 11301 D3 and 11301 D4, further from
sources of power frequency magnetic fields or to install
magnetic shielding. The power frequency magnetic
field should be me asured in the intended installation
location to assure that it is sufficiently low.