TOP CLAMP
FLOURESCENT
LIGHT
BOTTOM
CLAMP
NOTE:
Make a note of the wiring locations and the color of the wires before you
disconnect wires from the terminal block. Refer to your notes when you
reconnect the wires to the terminal block.
5. Remove the top four wires that connect the
The
TERMINAL BLOCK
6. Place one hand on the light and the other near the
Carefully, but with some force push the light to the left. The light will unclamp from
the
TOP CLAMP
7. Carefully pull the light forward and guide the unconnected wires out through the
RUBBER GROMMET
8. Carefully push and clamp the new light to the
9. Feed the unconnected wires back through the
the shaker.
10. Connect the top four wires to the
11. Lift up hinged front panel and replace the three screws. Use a Phillips (+) head
screwdriver to tighten the screws.
12. Replace and secure all shelving. Close shaker door and plug the
the electrical outlet.
New Brunswick Scientific
Figure 16: Light Replacement
is on the left, side wall of the shaker.
. Do the same for the
on the bottom of the shaker.
TERMINAL BLOCK
to the
LIGHT WIRES
TOP CLAMP
.
BOTTOM CLAMP
and
TOP
BOTTOM CLAMPS
RUBBER GROMMET
as removed.
.
TERMINAL BLOCK
(see Figure 16 above).
.
on the bottom of
into
POWER CORD
User's Guide