Maintenance

・ To clean the sensor, lightly wipe with a clean, soft cloth and
water diluted with a mild detergent, and then wipe with a dry,
Caution
soft cloth.
・ Do not wipe the sensor with organic solvent such as benzine.
・ Do not put oil such as grease on a wiper blade.

Maintenance (monthly)

Perform the following operations every month.
・ Clean the detection window and wiper blade using tap water.
・ Inspect the detection window for damage and/or degeneration.
・ Inspect the wiper blade for wear and/or deformation.
・ Inspect the wiper blade that it is tightly fixed.
・ Inspect the sensor cable for damage and/or degeneration.
・ Immerse the sensor in measurement water of the known concentration
and check that the correct value is displayed on the transmitter.

Periodic inspection (every 3 months)

Perform the following inspections every 3 months.
・ Terminal block screws of the transmitter are not rusted.

Replacement of consumables

・ Replace the wiper blade once a year as rough standard. In addition,
replace it when adequate wiping effect cannot be produced.
・ Replacement wiper blade and hanging bracket are included in the optional
maintenance kit (TSC-MK). For replacement, purchase TSC-MK.