10
Adjustment
Adjust and display the measurement values according to usage. In addition, adjust
and display the measurement values according to the environment where used.
To adjust, press
・ Pressing
・ If offset, span, or 2-point adjustment has already been
performed, be sure to initialize the adjustment before performing
the 2-point adjustment.
Caution
・ When "Display item (unit selection)" P.14 is changed, the
values of the offset adjustment, span adjustment, and 2-point
adjustment that have the display item are reset to the factory
default values.
Set again if necessary.
Offset adjustment
Set the offset coefficient to add to a
measurement value.
Display the measurement value with
offset-adjusted to any display value.
For 4-20 mA output ("7 4-20 mA Setting" P.16)
and alarm output ("Alarm output setting" P.18),
the adjusted values are applied.
Measurement item (unit)
[Display]
Turbidity (NTU, FNU, mg/L,
no unit)
[Turb.]
Turbidity ( 度 )
[Turb.]
Press
1
The measurement item selection display is displayed.
.
automatically retains the measurement value.
Setting range
-100.0 to 100.0 0.00
-10.0 to 10.0
.
Display value
after offset
adjustment
Measurement
value
Factory default setting
0.00