Turbidity Checker
TSC-10-D
Thank you very much for purchasing Turbidity Checker TSC-10-D.
All of this instruction manual must be read before operation of this
Turbidity Checker for safe and proper operation.
This instruction manual should be kept for future reference such as
maintenance.
The instruction manual is also available on the following website.
https://navi.optex.net/manual/05337/?lang=en
WATER it
Instruction
Manual

Advertisement

Table of Contents
loading

Summary of Contents for Optex TSC-10-D

  • Page 1 TSC-10-D Instruction Manual Thank you very much for purchasing Turbidity Checker TSC-10-D. All of this instruction manual must be read before operation of this Turbidity Checker for safe and proper operation. This instruction manual should be kept for future reference such as maintenance.
  • Page 2 The Contents of Packaging Sensor: 1 Cable protective cover: 1 Cable tying band: 5 Cleaning brush: 1 Detection window Instruction manual cleaning paper: 1 (this document) If there are any missing or faulty items, please contact our sales representative. Can be connected to SC-U1. Reference...
  • Page 3: Table Of Contents

    Table of Contents For Safe Use ................5 Component Name ..............7 Measuring Principle .............. 8 Installation ................9 Sensor installation ..............9 Extension of sensor cable ........... 12 Wiring ................... 13 Sensor Setting ..............14 Display item (unit selection) ..........14 Response time ..............
  • Page 4 11 Maintenance Setting ............30 Measurement value hold ............. 30 Hold timer ................30 Maintenance timer .............. 31 12 Cleaning ................32 Manual cleaning ..............32 Cleaning interval ..............33 13 Troubleshooting ..............34 Error code list ..............34 14 Maintenance ................. 36 Maintenance (monthly) ............
  • Page 5: For Safe Use

    ● Please thoroughly read "For Safe Use" before using this Turbidity Checker properly. ● Because these precautions are related to failure or malfunction, observe the precautions for use without fail. Do not use TSC-10-D except for measurement of water quality. " " denotes "Prohibited action", and "...
  • Page 6 To clean the sensor, first Do not wipe the sensor with wipe away lightly with a clean solvent such as Thinner. soft cloth damped by diluted It may cause failure. mild detergent solution and then wipe off moisture with a dry clean soft cloth. Avoid strong physical Do not apply oil such as shock to the sensor or do...
  • Page 7: Component Name

    Component Name Sensor cable Wiper blade Detection windows ・ Do not carry the sensor with the sensor cable. Caution ・ Do not touch the detection window. Clean the detection window if your hand touches it.
  • Page 8: Measuring Principle

    Measuring Principle Turbidity Checker TSC-10-D uses a method of 90 degrees scattered light. In the method, a light source of a sensor illuminates the surfaces of suspend- ed matter floating in the water, and the light is scattered by these surfaces is detected by a photo element installed at an angle of 90 degrees from the light source.
  • Page 9: Installation

    Installation Sensor installation Before installation, turn off the power switch of SC-U1 or disconnect the power cable from the supply source and wire the power cable after the installation is complete. " " denotes "Prohibited action", and " " denotes "Required action" . When installing the sensor, Do not install the sensor with make sure that the sensor...
  • Page 10 (TA-3) for fixing the sensor to the pipe is available as an option. For details, contact the dealer or visit our website: https://www.optex.co.jp/e/products/ environmental-sensor/ *Pipe: The 25A stainless steel pipe (outer diameter 34 mm) for piping to be prepared by the user.
  • Page 11 Wrap the cable protective cover on the sensor side of the sensor cable. Be sure to install the cable protective cover. The cable may be damaged and flooded during maintenance. Attach the end of the cable protective cover to a pipe with a cable tying band.
  • Page 12: Extension Of Sensor Cable

    Extension of sensor cable The standard length of the sensor cable is 9 m. To extend the cable, refer to the following. To extend the cable, connection using pull box is recommended. Nominal Max. cable length X sectional area 0.3 mm 20 m 0.5 mm 40 m...
  • Page 13: Wiring

    Wiring Use this product by connecting to Universal Transmitter SC-U1. For installation and wiring to SC-U1, refer to the "Universal Transmitter SC-U1 Installation Manual". Before wiring, turn off the power switch of SC-U1 or disconnect the power cable from the supply source and wire the power cable after all the other wiring is complete.
  • Page 14: Sensor Setting

    Sensor Setting The following settings are available from SC-U1. To start sensor setting, press Caution Pressing automatically retains the measurement value. Display item (unit selection) Set the unit of turbidity to show on the SC-U1 display. Main display For operation, refer to the "Universal Transmitter SC-U1 Operation Manual", "5.1 Sensor Setting"...
  • Page 15: Response Time

    ・ NTU: EPA-compliant unit. ・ FNU: Unit compliant with ISO7027. ・ mg/L: Unit selected for use as a simplified measure found by the cor- relation between the measurement value (e.g. floating suspended matter (SS) in the official method) and the value displayed on the Turbidity Checker.
  • Page 16: 4-20 Ma Setting

    4-20 mA Setting Set the 4-20 mA signal output setting from SC-U1. The factory default setting is OFF for 4-20 mA signal output. To set 4-20 mA output setting, press Caution Pressing automatically retains the measurement value. Output item selection Select an item to output by 4-20 mA signal.
  • Page 17: Upper/Lower Limit Of Signal Output Range

    Upper/lower limit of signal output range For the item by 4-20 mA signal output, set measurement value to output at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA). 20 mA Signal output 4 mA Measurement Measurement value at 4 mA value at 20 mA For operation, refer to the "Universal Transmitter SC-U1 Operation Manual", "5.3 4-20 mA Setting", "...
  • Page 18: Relay Setting

    Relay Setting Set the relay output setting from SC-U1. The factory default setting is OFF for relay output. To set the relay setting, press Caution Pressing automatically retains the measurement value. Alarm output setting Set the alarm to be output when the measurement value exceeds a certain value.
  • Page 19: Maintenance Output Setting

    Maintenance output setting Set a maintenance timer for any cycle (“Maintenance timer” P.31) to notify the user when the set period has elapsed. Use this for notification of sensor maintenance, replacement of consumables, and overhaul timing. Check notifications by the relay output and dedicated cloud server (SC-U1 and GW connected).
  • Page 20: Calibration

    Calibration Perform calibration before using the sensor or after cleaning. Distilled or deionized water is required. To calibrate SC-U1, press Caution Pressing automatically retains the measurement value. Zero calibration Immerse the sensor in distilled water and adjust the measurement value to the reference value (zero).
  • Page 21 Press The sensor selection display is displayed. Select a channel connected to TSC-10-D using and press Measurement Sensor The calibration menu is displayed. item model Sensor channel Select [ZERO Cal.] using , and press The calibration starts. During calibration a processing display is displayed.
  • Page 22 When the calibration is complete, check the result. good: The calibration is properly completed. good E4100 to 4199: Calibration failed. E4100 Try the steps again from 1. Pressing switches to the measurement value display.
  • Page 23: Adjustment

    Adjustment Adjust and display the measurement values according to usage. In addition, adjust and display the measurement values according to the environment where used. To adjust, press ・ Pressing automatically retains the measurement value. ・ If offset, span, or 2-point adjustment has already been performed, be sure to initialize the adjustment before performing the 2-point adjustment.
  • Page 24 Select [Turb. TSC-10-D] using , and press to switch pages. The adjustment menu is displayed. Sensor model Measurement item Sensor channel Select [Offset Adjustment] using , and press Enter the offset value. Select a digit by value by , and press When a value outside the set range is entered, a short beep sounds.
  • Page 25: Span Adjustment

    0.00 no unit) [Turb.] Turbidity ( 度 ) [Turb.] Press The measurement item selection display is displayed. Select [Turb. TSC-10-D] using , and press to switch pages. Sensor model The adjustment menu is displayed. Measurement item Sensor channel Select [Span Adjustment] using...
  • Page 26: 2-Point Adjustment

    Input a span adjustment value. Select a digit by value by , and press When a value outside the set range is entered, a short beep sounds. Pressing simultaneously while entering a value resets the value to the factory default setting. By pressing , the value returns to the one before changing it.
  • Page 27 0.00 [Turb.] [Point 2]: (1st point value + 2) to 100.0 100.0 Press The measurement item selection display is displayed. Select [Turb. TSC-10-D] using , and press to switch pages. Sensor model The adjustment menu is displayed. Measurement item Sensor channel...
  • Page 28 Enter the [Point 1] display, [Point 1] adjusted, [Point 2] display, and [Point 2] adjusted values in this order. Select a digit by and value by , and press . The value is set and the cursor moves to the next item. When a value outside the set range is entered, a short beep sounds.
  • Page 29: Adjustment Initialization

    Adjustment initialization Reset the adjustment value to the factory default value. Press The measurement item selection display is displayed. Select [Turb. TSC-10-D] using , and press to switch pages. Sensor model The adjustment menu is displayed. Measurement item Sensor channel...
  • Page 30: Maintenance Setting

    Maintenance Setting Set the maintenance setting from SC-U1. To perform maintenance, press Caution Pressing automatically retains the measurement value. Measurement value hold The latest measurement value is retained (held). This is used for sensor maintenance and cleaning. For operation, refer to the "Universal Transmitter SC-U1 Operation Manual", "8 Maintenance Setting"...
  • Page 31: Maintenance Timer

    Maintenance timer Set a maintenance timer for any cycle to notify the user when the set period has elapsed. This is used for notification of sensor maintenance, replacement of consumables, and overhaul timing. Check notifications by the relay output (“Maintenance output setting” P.19) and the dedicated cloud server (SC-U1 and GW connected).
  • Page 32: Cleaning

    Cleaning Set the sensor cleaning setting from SC-U1. To perform cleaning, press ・ Pressing automatically retains the measurement value. Caution ・ Do not press while a sensor with a wiper is in the air. This may cause the sensor to malfunction. Manual cleaning Activate the wiper manually.
  • Page 33: Cleaning Interval

    Cleaning interval Set a time interval to activate the wiper. This product is a sensor with wipers and it operates as following. Cleaning Cleaning not performed Cleaning Cleaning not performed 1 round 1 round trip of trip of Cleaning interval Cleaning interval wiper blade wiper blade...
  • Page 34: Troubleshooting

    Troubleshooting Error code list The following table describes error codes, causes and actions. If you find any problem other than the error codes, refer to “When in trouble” in the "Universal Transmitter SC-U1 Operation Manual". When the normal condition is not recovered after the countermeasure is taken, when any problem other than these occurs, or when requesting repairs, check the model and the serial No.
  • Page 35 Error code and details Cause Action E2000 The sensor cable Turn off power of the is not properly transmitter, check the Sensor processing/sensor wired. wiring, and be sure to wire properly. The sensor is out The sensor must be of order. repaired.
  • Page 36: Maintenance

    Maintenance ・ To clean the sensor, lightly wipe with a clean, soft cloth and water diluted with a mild detergent, and then wipe with a dry, Caution soft cloth. ・ Do not wipe the sensor with organic solvent such as benzine. ・...
  • Page 37: Calibration

    Calibration The sensor is designed to be able to measure stably for a prolonged period of time. In order to maintain the reliability of measurement, however, calibration should be carried out at least once a year. Long-term storage If the sensor is not used over a prolonged period, perform the following operations.
  • Page 38: Specifications

    Specifications Name Turbidity Checker Model TSC-10-D Measurement 0.00 - 1000.0 NTU/FNU (formazin standard solution) 0.00 - 100.0 度 (polystyrene standard solution) range Display 0.00 - 19.99: 0.01 resolution 20.0 - 1000.0: 0.1 Power supply 24 VDC±10% voltage Current Standard: 30 mA max.
  • Page 39: External Dimensions

    External Dimensions Ø48 (mm)
  • Page 40 ■ EU contact information ■ UK contact information https://navi.optex.net/cert/contact/ Whole document of the DOC can be referenced in the following website ; www.optex.net OPTEX CO., LTD. 5-8-12, Ogoto, Otsu, Shiga, 520-0101 Japan el.+81-77-579-8680 Fax.+81-77-579-8199 https://www.optex.co.jp/e/products/environmental-sensor/ 2021.08 59-3093-0...