Upper/Lower Limit Of Signal Output Range - Optex TSC-10-D Instruction Manual

Turbidity checker
Table of Contents

Advertisement

Upper/lower limit of signal output range

For the item by 4-20 mA signal output, set measurement value to output
at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA).
Signal output
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.3 4-20 mA Setting", " ◆ Signal output range lower limit"
and " ◆ Signal output range upper limit".
Output item
(unit)
[Display]
Turbidity
Lower limit:
(NTU, FNU,
0.00
mg/L, no unit)
Upper limit:
[Turb.]
1000.0
Turbidity ( 度 )
Lower limit:
[Turb.]
0.00
Upper limit:
100.0
The setting unit depends on the display resolution. (For display resolution,
refer to "15 Specifications" P.38)
20 mA
4 mA
Measurement
value at 4 mA
Signal output range
Factory de-
fault setting
0.00 to 1000.0
0.00 to 100.0
Measurement
value at 20 mA
Minimum setting range
Lower limit to
Upper limit
2
2
Upper limit - Lower
limit ≥
17