LOCalizer Reader and RFID Localization System User Manual Manufactured for: Health Beacons, Inc. 383 Lowell Rd. Concord, MA 01742 Phone: (978) 254-6500 ML-0004 Rev 06...
HB100 The LOCalizer Reader includes an integrated Loop Probe (A) that is used to locate and read Tags from the skin surface. It can also be used with the LOCalizer Surgical Probe attachment (D) to locate and read Tags within the...
Page 3
The LOCalizer Reader is designed only for use with the LOCalizer Surgical Probe and Tags. Use of the LOCalizer Reader adjacent to or stacked with other equipment should be avoided because it could result in improper functioning. If such use is necessary, this equipment and the other equipment should be Tag Applicator and Surgical Probe observed to verify that they are operating normally.
Handle all components in a manner that will prevent accidental contamination. presence of the Tag. Modulation of the eld by the Tag is detected by the LOCalizer Reader. Other devices that generate electromagnetic elds in this frequency region may interfere with the ability of the LOCalizer to sense the location of the Tag.
Page 5
LOCALIZER READER CONTROLS The LOCalizer Reader contains the display and controls for the System and are located on the front face of the Control/ instrument. The LOCalizer Reader allows the user to adjust the system’s settings, and produces signals in the form...
Page 6
NOTE: The LOCalizer Reader may turn ON with a critically low battery, but will not continue operation normally. Use the Loop Probe on the LOCalizer Reader to identify the general location of the Tag from the surface of the breast. The sound level and pitch increases as the LOCalizer Reader approaches the Tag. The scale shows the approximate distance to the nearest point of the Tag in millimeters.
The Tag ID is displayed only when the Loop probe is within 40mm of the Tag Tag ID is not displayed or display these may degrade the case. Do NOT submerge the LOCalizer Reader in liquid. and the Surgical Probe is within 20mm. If within these limits, there may be an shows “...
50kPa - 106kPa Pressure *For best service life of AA Alkaline Cells used in LOCalizer Reader, storage temperature should be 5° - 30°C. To avoid potential damage from cells leaking, remove battery if the LOCalizer Reader will not be used within 2 months.
6V rms, on ISM bands Basic test standard LOCalizer Reader at the end of the LOCalizer Reader’s useable life per Waste of Electrical and Electronic on SIP/SOP port only Equipment (WEEE) directive [Directive 2002/96/EC]. Dispose of Tag Applicator, Surgical Probe, and Instrument...