General Warnings and Cautions

General Warnings and Cautions
The following list of warnings and cautions apply to
general operation and maintenance of the Fabius Tiro.
Warnings and cautions about installing and operating
specific parts appear with those topics.
Warning statements give important information
that, if ignored, could lead directly to a patient's
or operator's injury.
Caution statements give important information
that, if ignored, could lead directly to equipment
damage and, indirectly, to a patient's injury.
WARNING !
Any person involved with the setup, operation, or
maintenance of the Fabius Tiro anesthesia system
must be thoroughly familiar with this instruction
manual.
WARNING !
This anesthesia system will not respond automatically
to certain changes in patient condition, operator error,
or failure of components. The system is designed to
be operated under the constant surveillance and
control of a qualified operator.
WARNING !
No third-party components shall be attached to the
anesthesia machine, ventilator, or breathing system
(except for certain approved exceptions). For more
information, contact DrägerService.
WARNING !
Each institution and user has a duty to independently
assess, based on its, his, or her unique
circumstances, what components to include in an
anesthesia system. However, Dräger, in the interest of
patient safety, strongly recommends the use of an
oxygen analyzer, pressure monitor, volume monitor,
and end-tidal CO
monitor in the breathing circuit at all
2
times.
Fabius Tiro Operator's Manual
‫.ﻣراﺟﻌﮫ ﻓرﻣﺎﯾﯾد‬WWW.ELM-FAN.IR‫ﺑرای داﻧﻠود ﻣوارد ﺑﯾﺷﺗر ﺑﮫ ﺳﺎﯾت‬
To download more items, go to ELM-FAN.IR
Chapter 1 – Introduction
WARNING !
Exercise special care so that the machine does not tip
when moving up or down inclines, around corners,
and across thresholds (for example, in door frames
and elevators). Do not attempt to pull the machine
over any hoses, cords, or other obstacles on the floor.
WARNING !
Apply the caster brakes when the anesthesia machine
is in use.
WARNING !
Do not place third-party components on top of the
Fabius Tiro monitor housing. Only components
approved by Dräger may be used.
WARNING !
Route all lines/cables away from the APL valve to
prevent interference with the APL valve adjustment
knob. Lines/cables caught underneath the APL valve
adjustment knob could interfere with proper
functioning of this valve.
11