Titan atrato Instruction Manual page 42

Ultrasonic flowmeter range
Hide thumbs Also See for atrato:
Table of Contents

Advertisement

Titan Interface Software Opens but
shows not connected to meter (no
spinning bars in bottom left of screen)
Titan
Interface
Software
showing all Tabs (User Settings; Flow
Details; Advanced; Datalogging)
Moisture in Display Window
Leak from end of the body of meter
42
Titan Enterprises Ltd ©2021
®
Settings incorrect for Range
Two windows of the
Interface Software are open
and second window is
masking connected window
Damage to USB
Connection
is
not
Software is placing the tab
window outside the monitor
area
Condensation or water in
the flowmeter
Poor seal on inlet/outlet
Housing damage on
polymer end caps leading
to internals leaking
Use Titan Interface software
to range the analog output
flow min and max points
Either drag window out of
the way or close second
window using task bar
Return to Titan Enterprises
for repair
Go to View on the top bar
menu and click Restore
Defaults. This restores the
Viewing defaults and re
docks all tabs to the main
window
Dry unit before use
Extensive water
contamination will cause
multiple errors on the device
and will require repair at
Titan Enterprises
Correct seal
If damage is seen, return
unit to Titan for repair and
recalibration
Atrato_Instruction_Manual_0322