Single Flow Alarm:
NOTE:
Alarms use Filtered Flow Values!
Checking this option opens up input boxes for
Threshold, Latching
Threshold
Set the
transistor to operate.
The logic of operation can be adjusted with the
Invert
check box.
Latching: This gives the user three options of None;
High Threshold and Low Threshold.
NONE will cause the alarm to be triggered when the
threshold flow is met. The alarm state will then reset
in accordance with the user set Hysteresis.
HIGH THRESHOLD will cause an alarm signal to
Latch on when flow is seen above the set value.
LOW THRESHOLD will cause an alarm signal to
Latch on when flow is seen below the set value.
21
Titan Enterprises Ltd ©2021
®
Invert.
and
to the flow value you require the
Atrato_Instruction_Manual_0322