ECscan10 Pocket Conductivity Tester USER MANUAL Bante Instruments Inc.
Page 2
If you do not press any key within 8 minutes, the tester will switch off automatically to conserve energy. turn clockwise until tight. Conductivity Calibration The ECscan10 conductivity tester allows 1 point calibration. For better accuracy, we recommend to calibrate the tester regularly. The following table shows acceptable standard solutions for each tester. Keypad...
Page 3
ECscan10 Measurement Conductivity Measurement Rinse the electrode with distilled water, place the electrode into the sample solution and stir gently, make sure that no air bubbles on the sensor surface. Wait for the measurement to stabilize and record the 2. Press and hold the Cal key to enter the calibration mode, the reading.
Page 4
ECscan10 Replacing the Electrode Conductivity If the tester fails to calibrate or gives fluctuating readings, you should ECscan10 L: 1.0 to 199.9 μS/cm consider replacing the electrode. Range ECscan10 M: 10 to 1999 μS/cm 1. Twist the electrode collar counter clockwise, pull the electrode away from the tester.