Advertisement

Quick Links

Operating instructions
SIMPLEX
Vacuum systems
VD 0025 G – 0063 G – 0100 G
Dr.-Ing. K. Busch GmbH
Schauinslandstraße 1
79689 Maulburg
Germany
0870169324/-0006_en / Translation of the original operating instructions / Subject to change without notice
16.11.2021

Advertisement

Table of Contents
loading

Summary of Contents for BUSCH SIMPLEX

  • Page 1 Operating instructions SIMPLEX Vacuum systems VD 0025 G – 0063 G – 0100 G Dr.-Ing. K. Busch GmbH Schauinslandstraße 1 79689 Maulburg Germany 0870169324/-0006_en / Translation of the original operating instructions / Subject to change without notice 16.11.2021...
  • Page 2               0870169324_VD 0025‐0100G_‐0006_IM_en 2 / 46   ...
  • Page 3: Table Of Contents

    1 | Contents Contents 1.1 Table of contents Contents ..................... 3 Table of contents ......................... 3 Register of tables ......................... 5 Register of illustrations ........................ 5 Safety ....................6 Safety devices ..........................6 Emergency information ....................... 6 Product description ................7 View vacuum system VD ......................
  • Page 4 1 | Contents Troubleshooting ................27 General faults ..........................27 Failure table and measures ....................... 28 Spare parts and accessories ...............30 10.1 Spare parts ..........................30 Repair ....................31 Decommissioning and disposal ............32 12.1 Stop operating vacuum system ....................32 12.2 Disassembly and disposal ......................32 Dimensions sheets ................33 Connection values of the vacuum system VD ........39 14.1...
  • Page 5: Register Of Tables

    1 | Contents 1.2 Register of tables Tab. 1: Connections vacuum system VD 0025 G AAA TAXX ........33 Tab. 2: Connections vacuum system VD 0025 G AAA TAAX (mobile) ......34 Tab. 3: Connections vacuum system VD 0063 G AAA TCXX ........35 Tab.
  • Page 6: Safety

        2 | Safety  Safety  Before commissioning of the vacuum system, read these operating instructions with care.  Please contact your contact person from Busch if there are any questions.  Keep the operating instructions so that you can use them for reference at a later time if nec‐ essary.  These operating instructions remain valid as long as the customer does not make any changes  to the product.  The vacuum system is intended for industrial use. It must only be operated by technically  trained specialists.  Always wear personal protective equipment in accordance with local regulations.  The vacuum system has been designed and produced according to state‐of‐the‐art methods.  Nevertheless, a residual risk remains in operation. Potential dangers are highlighted in these  operating instructions. Safety and warning notes are marked as follows, with the words DAN‐ GER, WARNING, CAUTION, ATTENTION and NOTE:    DANGER  ... Indicates a threatening hazard. Failure to observe safety instructions will result in death or  serious injury.    WARNING  ... Indicates a potential hazard. Failure to observe safety instructions may result in death or se‐ rious injury.    CAUTION  ... Indicates a potential hazard. Failure to observe safety instructions may result in slight inju‐ ry.    ATTENTION!  ... Indicates a potential hazard. Failure to observe safety instructions may result in property  damage.    NOTICE  ... Indicates helpful advice and recommendations as well as information for efficient and  smooth operation.  2.1 Safety devices  The motor of the rotary vane vacuum pump is safeguarded by a motor protection switch. If an ...
  • Page 7: Product Description

          3 | Product description  Product description  3.1 View vacuum system VD              IN  Gas inlet  CC  Switch and control cabinet  OUT  Gas outlet  AF  Air filter  INO  Gas inlet (optional)  VV  Venting valve  RVP  Rotary vane vacuum pump RA  CPG  Contact vacuum gauge  NP  Name plate  VST  Vacuum vessel  Fig. 1:  View vacuum system VD  0870169324_VD 0025‐0100G_‐0006_IM_en  7 / 46  ...
  • Page 8: View Of The Rotary Vane Vacuum Pump R5 Ra

    DA  Rotation direction arrow motor  OFP  Oil fill plug  EF  Exhaust filter  OSG  Oil sight glass  NP  Name plate  ODP  Oil drain plug  OF  Oil filter  EB  Eye bolt  AF  Axial fan  GB  Gas ballast valve  OS  Oil separator  Fig. 2:  View rotary vane vacuum pump type R5 RA (0025 F)  3.3 Setup  The vacuum system SIMPLEX VD comprises of a single‐stage rotary vane vacuum pump  type R5 RA. The vacuum pump is mounted on a vacuum vessel.  The intake‐side air filter and the strainer installed in the suction flange prevent the ingression  of dirt particles in the vacuum pump.  The gas ballast valve is used for the addition of a limited quantity of ambient air to the pro‐ cess gas, to counteract the condensation of vapor in the vacuum pump.  In the vacuum vessel is a contact vacuum gauge installed for the 2‐point control which dis‐ plays the pressure in the vacuum vessel.  The vacuum system is completely piped and the electrical components are wired on a switch  and control cabinet.  The vacuum system is also available as a mobile version on a push bar cart with four rolls.  0870169324_VD 0025‐0100G_‐0006_IM_en 8 / 46   ...
  • Page 9: Function Principle

    3.4 Function principle  3.4.1 Vacuum system VD  Gas delivery from the vacuum vessel is effected by the single‐stage rotary vane vacuum pump.  Pressure gas is exhausted against the atmosphere.  The target values are set to the desired switch‐on and switch‐off values on the contact vacu‐ um gauge on the vacuum vessel. When the lower switching point is reached, the vacuum  pump is switched off. When the vacuum in the vacuum vessel is used up, the pressure in the  vessel increases and when the upper switching point is reached, the vacuum pump is switched  on again and evacuates the vacuum vessel.  3.4.2 Rotary vane vacuum pump R5 RA    Fig. 3:  Operating principle of the rotary vane vacuum pump R5 RA  The vacuum pump R5 RA works according to the rotary vane principle.  The oil seals the spaces, lubricates the sliders and discharges the compression heat. The oil fil‐ ter cleans the circulating oil.  Then the exhaust filters separate the oil from the discharged gas.  3.4.3 Gas ballast valve  The vacuum pump is equipped with a gas ballast valve. The gas ballast valve is used for the  addition of a limited quantity of ambient air to the process gas, to counteract the condensa‐ tion of vapor in the vacuum pump. The gas ballast valve reduces the final pressure of the vac‐ uum pump, see “Technical data” in chap. 15.  3.5 Intended use  The vacuum system was designed for conveying air and other dry, non‐aggressive, non‐toxic  and non‐explosive gases.  Conveying any other media leads to increased thermal and/or mechanical stress of the vacu‐ um system and is only permitted in coordination with Busch.  The vacuum system is designed for operation in a non‐hazardous area. The vacuum system  can be operated continuously at final pressure and is suitable for continuous operation. The  permitted ambiance conditions can be found in the technical data (chap. 15).  The vacuum system is designed for indoor use; for outdoor installation, contact Busch to  make special arrangements if necessary.  0870169324_VD 0025‐0100G_‐0006_IM_en  9 / 46  ...
  • Page 10: Transport

        4 | Transport  Transport    WARNING  Danger of severe injury!  Suspended load.  • Never walk, stand or work below suspended loads.    WARNING  Hazard from vacuum system falling or tipping over!  The weight of the vacuum system can kill a person or cause severe crushing.  • Use a pallet appropriate to the weight and size of the system on which it can be moved by  a forklift truck. Or lift the vacuum system using slings and suitable lifting gears with a fork‐ lift or lift truck to move it or remove the pallet.  • Take care to avoid strain on the pipe work and vacuum pumps while placing the slings.  • Observe the center of gravity and the lifting points; these are indicated in the scale draw‐ ings, see chap. 13.    ATTENTION!  Damage to the vacuum system!  The vacuum system is already filled with oil at delivery.  • Drain the oil before transport if transport in horizontal orientation is not possible.  The vacuum system is packed in a wooden crate. It protects the system from damage during  transport.  The vacuum system is packed in a wooden structure and can be moved using a forklift.   Unpack the vacuum system as near to the installation site as possible.   Check scope of delivery for completeness.   Check the vacuum system for transport damage.   Dispose of packing material as required by current regulations.   ...
  • Page 11       4 | Transport      WARNING  Danger of severe injury!  Lift the vacuum system by devices of the individual components.  • Do not lift the vacuum system by devices of the individual components,   e.g. the vacuum pump, motor, etc.   • Only lift the vacuum system as presented.      Fig. 4:  Transport of the vacuum system  Screw four M10 eye bolts (not included in the scope of delivery) into the holes on the feet of  the vacuum vessel (see fig. 4). Pass slings through the openings of the eye bolts and thus lift  the vacuum system with a crane or a forklift.                      0870169324_VD 0025‐0100G_‐0006_IM_en  11 / 46  ...
  • Page 12     4 | Transport    The vacuum pump can be lifted by the eye bolts.    WARNING  Danger of severe injury!  Suspended load.  • Never walk, stand and work below suspended loads!   • The eye bolts (EB) must be in perfect condition and completely screwed into the machine  and hand‐tightened!  • Do not lift the vacuum pump on an eye bolt of the motor. Lift the vacuum pump only as  shown.      ATTENTION!  Damage to the vacuum system!  The vacuum system is already filled with oil at delivery.  Tilting a vacuum pump that is already filled with oil can cause large amounts of oil to enter  the cylinder. If the vacuum pump is started while there are excessive amounts of oil in the  cylinder, this will damage the sliders, resulting in a total damage of the vacuum pump.  • Drain the oil before transport if transport in horizontal orientation is not possible.    Machine weight: see technical data or nameplate (NP)   Fig. 5:  Transport of the vacuum pump on the eye bolts      0870169324_VD 0025‐0100G_‐0006_IM_en 12 / 46   ...
  • Page 13: Storage

          5 | Storage  Storage    ATTENTION!  Danger of damage to the vacuum pump drive!  Long storage periods may cause capacitors in the drive to be weakened by electro‐chemical  condensation. In the most detrimental case, this may cause short‐circuit and thereby destruc‐ tion of the drive.  • The vacuum pump should therefore be connected to the power supply for 30 minutes eve‐ ry 18 months.  Proceed as follows for storage:   Close all openings with the protective caps included in the scope of delivery (penetration of  dirt and water is prevented)  If storage for more than 3 months is intended:   Secure loose cables   Drain all process and operating media   Clean and dry the vacuum system (Prior to storing make absolutely sure that all parts are  clean, drained and dry)   Where necessary use oil for conservation   Wrap the vacuum system in corrosion‐inhibiting film.   Store the vacuum system in a protected, dry and dust‐free room at a temperature between  0 and 40 °C.  0870169324_VD 0025‐0100G_‐0006_IM_en  13 / 46  ...
  • Page 14: Installation

        6 | Installation  Installation  6.1 Installation    Fig. 6:  Installation environment   Ensure that the vacuum system is set up horizontally (deviation 1° max.) and anchor it in the  ground with four bolts if necessary.   Technical data must be complied with.   The ambiance conditions must meet the protection class of the vacuum pump.   The installation site must be vented so that sufficient cooling of the vacuum system is en‐ sured.   Ensure that the ventilation openings (inlets and outlets) are not covered and that the cool‐ ing air can flow unhindered.   Sufficient space for maintenance work must be ensured.   Visibility of the oil sight glass (OSG, fig. 2) must be ensured at all times.   Check the oil level and top up oil if necessary (for more information, see topping up with oil  (chap. 7.1.1)).   Ensure that all covers, safety devices, etc. are installed.  0870169324_VD 0025‐0100G_‐0006_IM_en 14 / 46   ...
  • Page 15: Connection Lines/Pipes

    • Remove all foreign matter (welding beads, filings, etc.) from the pipelines! This may be  done by flushing or blowing through pipelines.  • The client must ensure that the pipework at the inlet is clean.  .  ATTENTION!  Vacuum system can be damaged by tensile or compressive stresses on pipeline connec‐ tions!  • If stresses could occur, use compensators to connect pipelines.    ATTENTION!  Condensate damages vacuum pumps!  • Lay the piping on the intake and pressure sides at an angle to prevent accumulated con‐ densation from entering the vacuum pump.  The line cross‐section of the connection lines must have at least the same cross‐section as the  connections of the vacuum system across the entire length.  In case of very long connection lines, it is recommended to use lines with larger cross‐sections  to avoid loss of efficiency. Please contact your contact person from Busch.  6.2.1 Gas inlet    WARNING  Danger of severe injury!  Open gas inlet!   • Never insert your hand or fingers into the gas inlet!  Connect the suction side pipeline to the gas inlet of the vacuum system (IN, fig. 1).    NOTICE  Before initial start‐up check the pipelines to the vacuum system for leaks. Rectify any leaks.  Dimensions see scale drawings in the appendix.        0870169324_VD 0025‐0100G_‐0006_IM_en  15 / 46...
  • Page 16: Gas Outlet

        6 | Installation  6.2.2 Gas outlet    CAUTION  Health risk!  The discharged gas contains small amounts of oil.  Ensure sufficient ventilation in the installation room when the air is routed into rooms where  there are people.   Connect the gas outlet line to both gas outlet openings of the vacuum pump if required.  Dimensions, see scale drawing in the appendix; connection size of thread: G 1 ¼“.   Ensure that the discharged gas can flow off unhindered. Never close the gas outlet line, do  not throttle it and do not use it as a compressed air source.  If the air taken in is not discharged to the environment in the direct proximity of the vacuum  system, observe the following:   Place the gas outlet line dropping towards the vacuum system or install a liquid separator or  a siphon with a drain valve so that no liquid can flow back into the vacuum system.      Fig. 7:  Gas outlet        0870169324_VD 0025‐0100G_‐0006_IM_en 16 / 46   ...
  • Page 17: Electrical Connection

    Incorrect rotation of drive motors can seriously damage the vacuum system!  Switch vacuum pumps briefly on and off again to check the rotational direction of the motor.  The rotational direction is marked by a direction of rotation arrow on the motor.  If the direction is wrong, reverse two connection phase poles.  6.3.1 Connection of the power supply    ATTENTION!  Danger of damage to the vacuum system.  Wrong connection.  Wire the vacuum system according to the circuit diagram in the switch and control   cabinet.  Procedure:   Ensure that the power supply for the vacuum system corresponds to the specifications in  the circuit diagram.   Ensure that the vacuum system is not impaired by electrical or electromagnetic impulses of  the power supply. Contact Busch if necessary.   Connect the switch and control cabinet to the power supply (connection values, see  chap. 14).      ATTENTION!  Danger of damage to the motor  Wrong rotating direction.  • Operation in the wrong rotating direction may damage the vacuum system after a very  brief time. Ensure that the rotating direction is correct before commissioning.   Use the glued‐on/cast‐in arrow to determine the intended rotating direction.   Switch on the vacuum pump for a fraction of a second.   Observe the fan wheel and determine the rotating direction just before standstill.  To change the rotating direction:   Swap any two phases of the power supply.  0870169324_VD 0025‐0100G_‐0006_IM_en ...
  • Page 18: Start-Up

        7 | Start‐up  Start‐up  7.1 Indication and control elements    DANGER  Danger from electric shock!  Electric shock will cause death or serious injury.  • Before commissioning, ensure that all electrical lines are covered and that the terminal box  is closed!    CAUTION  Burn hazard!  The surface of the vacuum pump may reach operating temperatures in excess of 70 °C in  operation.  Do not touch the vacuum pump during and directly after operation.    ATTENTION!  Insufficient familiarity with the indicators and operating controls may result in damage to  the vacuum system.  Wrong operation.  • Operating staff must be familiar with the indicating and operating elements.   ATTENTION!  Operation of the vacuum pump without oil will cause severe damage to it quickly.  • The vacuum system is already filled with oil at delivery. The oil level must be checked and,  if necessary, topped up before commissioning.  7.1.1 Topping up with oil  For oil type and oil quantity, see Technical data (chap. 15) and Oil (chap. 16).  1 x sealing ring, part no.:  0486 000 590  Check oil level    0870169324_VD 0025‐0100G_‐0006_IM_en 18 / 46   ...
  • Page 19: Display And Control Elements At The Switching And Control Cabinet

          7 | Start‐up  Fig. 8:  Topping up oil  7.1.2 Display and control elements at the switching and control cab‐ inet    Fig. 9:  Switch and control cabinet   Pushbutton with indicator lamp (101S1)  Pushbutton for switching the vacuum system on and off  Indicator lamp lights up during operation   Indicator lamp of the collective failure (101H1)  Indicator lamp lights up red when there is an error   Operating hours counter (101T1)  Displays the operating hours   Main switch (100Q1)  Switch for switching the supply voltage on and off  7.1.3 Indicators and control elements on the vacuum system  The indicating devices are as follows:   Contact vacuum gauge (CPG, fig. 1)  Indicates the pressure in the vacuum vessel.  The switch‐on and switch‐off value (pressure range) is set on the contact vacuum gauge.   Venting valve (VV, fig. 1)  The vacuum system is vented to atmospheric pressure via this valve for maintenance work.   Oil sight glass (OSG, fig. 2) at the vacuum pump      0870169324_VD 0025‐0100G_‐0006_IM_en  19 / 46  ...
  • Page 20: Setting The Target Value Needle On The Contact Vacuum Gauge

        7 | Start‐up  7.1.4 Setting the target value needle on the contact vacuum gauge  The target values are set via the adjustment lock in the sight disc using the adjustment key  (enclosed, can be found on the side of the cable socket). To adjust the set pointer, place the  adjustment key at the adjustment lock, press it in and set the required set point.    Set pointers Removable  adjustment key  Adjustment lock   Fig. 10:  Contact vacuum gauge  The set pointers of the limit switches can be adjusted freely throughout the scale range. For  reasons of switching accuracy, switching safety and service life of the mechanical measuring  systems, the switching points should be between 10 % and 90 % of the measuring range.  Set the set pointers to desired activation and shutdown pressures. The clearance between the  two set pointers is approx. 10% of the scale range.  When the lower switching point is reached, the vacuum pump is switched off and correspond‐ ing, when the upper switching point is reached, again switched on.  Factory settings of the switching points:   lower switching point: ‐ 0.9 bar (90 %)   upper switching point: ‐ 0.8 bar (80 %)      NOTICE  The recommended minimum clearance between two contacts is 10% of the measuring span.  The switch hysteresis is 2 … 5%.          0870169324_VD 0025‐0100G_‐0006_IM_en 20 / 46   ...
  • Page 21: Operation

    Starting the Vacuum System   Set the main switch (100Q1, fig. 9) to “ON”.   Press the pushbutton (101S1, fig. 9) on the switch and control cabinet.   The vacuum pump switches on and off depending on the pressure. The vacuum system is in  operation.  Stopping the vacuum system   Press the pushbutton (101S1, fig. 9) on the switch and control cabinet.   The motor switches off.   Set the main switch (100Q1, fig. 9) to “OFF”.  7.3 Conveying condensing vapors  A certain quantity of water vapor within the gas flow is tolerated.  For information, see technical data (chap. 15).  Contact Busch to get information on transporting other vapors.  Note the following when conveying condensing vapors:   Start the vacuum system, all vacuum pumps must be running.   Close the shut‐off valve on the suction side (not part of the scope of delivery)   Allow the vacuum pumps to warm up (approx. 30 min.) so that no condensate can form in  the suction chamber.   Open the shut‐off valve on the suction side and start your process   Close the shut‐off valve on the suction side after the end of the process   The vacuum pumps must now run for approx. 30 min. to remove any moisture that may be  present.   Switch off the vacuum system.  0870169324_VD 0025‐0100G_‐0006_IM_en  21 / 46...
  • Page 22: Maintenance

      WARNING  Danger from vacuum pump contaminated with hazardous material!  There is a danger of poisoning!  • Vent the vacuum pumps before they are maintained  8.1 Maintenance plan  The maintenance intervals strongly depend on the individual operating conditions. The inter‐ vals specified below are to be considered reference values and should be individually short‐ ened or extended. Especially at high stress, e.g. in case of high dust load in the environment or  the process gas, or in case of other contamination or ingression of process material, it may be  necessary to shorten the maintenance intervals a lot.  Interval  Maintenance task  Rotary vane vacuum pump R5 RA  • Check the oil level, see chap. 8.2.  daily  • Check the vacuum pump for escaping oil. Have the  vacuum pump repaired by Busch if there are any leaks.  Monthly  • Check the filter insert in the suction filter and replace  if necessary, see chap. 8.5.  • Clean the vacuum pump from dust and contamination.  six‐monthly  • Clean the filter of the gas ballast valve (GB, fig. 2)  • Replace the oil*, the oil filter* (OF, fig. 2, chap. 8.3)  Normal use: after max.  4,000 operating hours or  and exhaust filters (EF, fig. 2, chap. 8.4).  after one year at the latest      *Note: Maintenance interval for synthetic oil.  Highly demanding purpose: ...
  • Page 23: Check Oil Level

          8 | Maintenance  8.2 Check oil level  Proceed as follows:   Switch off the vacuum system.   Wait 1 minute after switching off the vacuum pumps, before checking the oil level.      Top up with oil if necessary.    8.3 Oil and oil filter change    ATTENTION!  Operation of the vacuum pump without oil will cause severe damage to it quickly.  • Only use oils approved by Busch.   For information to the oil type and oil volume, see the technical data   (chap. 15) and oil (chap. 16).  Follow the figures  1 x sealing ring, part no.:  Drain pan  0486 000 505      Fig. 11:  Drain oil        0870169324_VD 0025‐0100G_‐0006_IM_en  23 / 46  ...
  • Page 24     8 | Maintenance  Original spare parts from Busch  1 x oil filter (OF), part no.: 0531 000 002  Oil filter wrench      Fig. 12:  Replacing the oil filter    1 x sealing ring, part no.:  0486 000 590  Check oil level    Fig. 13:  Filling in oil  0870169324_VD 0025‐0100G_‐0006_IM_en 24 / 46   ...
  • Page 25: Replacing The Exhaust Filters

          8 | Maintenance  8.4 Replacing the exhaust filters  Follow the figures  RA 0025/0040 F: 1 x exhaust filter (EF)  10 mm wrench  RA 0063/0100 F: 2 x exhaust filter (EF)    Fig. 14:  Remove exhaust filter    Original spare parts from Busch  RA 0025/0040 F: 1 x exhaust filter (EF)  part no.: 0532 140 156  RA 0063/0100 F: 2 x exhaust filter (EF)  part no.: 0532 140 157  O‐ring  10 mm wrench  RA 0025/0040 F: 1 x gasket  RA 0063/0100 F: 2 x gasket  Part no.: 0480 000 112      Fig. 15:  Insert and tighten the new exhaust filter  0870169324_VD 0025‐0100G_‐0006_IM_en  25 / 46  ...
  • Page 26: Changing The Air Filter Insert

        8 | Maintenance  8.5 Changing the air filter insert      Fig. 16:  Changing the air filter insert  Proceed as follows:   Open the quick clamps and remove the filter insert   Insert a new filter.  Original spare part from Busch: Parts no. 0532 000 003      0870169324_VD 0025‐0100G_‐0006_IM_en 26 / 46   ...
  • Page 27: Troubleshooting

          9 | Troubleshooting  Troubleshooting    DANGER  Electric shock danger.  Live wires.  • Electrical installation work must only be performed by qualified specialists.    WARNING  Hazard from running vacuum pump!  There is a severe threat for life and limb depending on the repair and maintenance work to  be carried out if the vacuum pump is running.  Only carry out any repair or maintenance work if:   the vacuum pump is not running   the vacuum pump is disconnected from the electrical supply and secured against uninten‐ tional start‐up   vacuum system and pipes are not under pressure,   hot surfaces are cooled down,  Disconnect the vacuum system from the process and make sure that there is ambient pres‐ sure in the vacuum system! Procedure: Close the shut‐off device on the inlet side (not includ‐ ed in the scope of delivery) and slowly open the venting valve (AF, fig. 1).  9.1 General faults  The figure shows the components that are relevant for troubleshooting.      FV  Float valve  EF  Exhaust filter  IS  Intake screen  GB ...
  • Page 28: Failure Table And Measures

    • Replace the air filter insert.  The air filter insert   (fig. 18) is clogged  • Close the venting valve.  The venting valve (VV,  fig. 1) is open.  • Top up oil.  The oil level is too low  • Have the vacuum pump  Inner components are  worn or damaged  repaired   (Busch Service)   • Locate and repair leakage.  Leak in piping system  • Have the vacuum pump  High noise development  The bearings are defec‐ during operation of the  tive.  repaired (Busch Service).  vacuum pump  • Replace the coupling (CPL,  The coupling (CPL, fig. 17)  is worn.  fig. 17).  • Have the vacuum pump  The sliders are stuck. ...
  • Page 29 The exhaust filters (EF,  from the gas outlet of the  fig. 2 and 17) are partially  ters (EF, fig. 2 and 17).  vacuum pump.  clogged.  • Determine the proper  Exhaust filters (EF, fig. 2  and 17) and o‐rings are not  position of the exhaust fil‐ installed properly.  ters (EF, fig. 2 and 17) and  o‐rings.  • Check the float valve.  The float valve (FV, fig. 17)  • Let the vacuum pump re‐ does not work properly.  paired by Busch.  • Flush the vacuum pump.  The oil is black.  The periods between the  oil changes are too long.  Please contact your con‐ tact person from Busch  for this.  • Replace air filter.  The air filter (AF, fig. 1) is  defective.  • See "High noise develop‐ There is a high heat devel‐ opment during operation  ment during operation of  of the vacuum pump.  the vacuum pump". ...
  • Page 30: Spare Parts And Accessories

        10 | Spare parts and accessories  10 Spare parts and accessories  10.1 Spare parts    ATTENTION!  There is a risk of premature failure of the vacuum system.  Loss of efficiency when using non‐original (non‐Busch) spare parts.  • We recommend using only original spare parts and consumables from Busch to ensure  proper functioning of the vacuum system and to fulfill all warranty and guarantee‐related  requirements.  Spare part  Description  Part number  Contains all spare parts needed for   0992 S01 700  maintenance, comprising:  1 x maintenance kit RA 0025 D  0992 101 463  Maintenance kit  1 x air filter insert (paper) for suc‐ VD 0025 G  0532 000 003  tion filter  1 x vacuum pump oil VM 100,  0831 000 060  package size 1 liter  Contains all spare parts needed for   0992 S01 701  maintenance, comprising:  1 x maintenance kit RA 0063 F,   0992 106 214  RA 0100 F  Maintenance kit  VD 0063 / 0100 G  1 x air filter insert (paper) for suc‐...
  • Page 31: Repair

          11 | Repair  11 Repair  Observe the following notes if the vacuum pump is returned to Busch.  Proceed as follows when the vacuum pump was used to transport gas contaminated with for‐ eign substances hazardous to health:   Decontaminate the vacuum pump and indicate the contamination status based on a "Con‐ formation declaration".  Busch accepts only vacuum pumps that include a completed and legally signed "Contamina‐ tion declaration".  The form can be downloaded from www.buschvacuum.com.    0870169324_VD 0025‐0100G_‐0006_IM_en  31 / 46  ...
  • Page 32: Decommissioning And Disposal

        12 | Decommissioning and disposal  12 Decommissioning and disposal  12.1 Stop operating vacuum system   Disconnect the vacuum system from the power supply.   Vent all connected pipes to atmospheric pressure.   Disconnect all connections.  If the vacuum system has to be stocked, observe following:   For more information, see storage (chap. 5).  12.2 Disassembly and disposal   Drain the oil.   Remove the exhaust filters.   Remove the oil filter.   Remove all electrical components.   Separate the hazardous waste from the vacuum pump.   Dispose of hazardous waste according to the applicable legal provisions.   Dispose of the vacuum system as waste metal and electronic scrap.    0870169324_VD 0025‐0100G_‐0006_IM_en 32 / 46   ...
  • Page 33: Dimensions Sheets

          13 | Dimensions sheets  13 Dimensions sheets  Dimensions vacuum system VD 0025 G AAA TAXX    Fig. 18:  Dimensions vacuum system VD 0025 G AAA TAXX    Center of  Lifting point      gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 1:  Connections vacuum system VD 0025 G AAA TAXX  0870169324_VD 0025‐0100G_‐0006_IM_en  33 / 46  ...
  • Page 34     13 | Dimensions sheets  Dimensions vacuum system VD 0025 G AAA TAAX (mobile)    Fig. 19:  Dimensions vacuum system VD 0025 G AAA TAAX (mobile)    Center of        gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 2:  Connections vacuum system VD 0025 G AAA TAAX (mobile)      0870169324_VD 0025‐0100G_‐0006_IM_en 34 / 46   ...
  • Page 35       13 | Dimensions sheets  Dimensions vacuum system VD 0063 G AAA TCXX    Fig. 20:  Dimensions vacuum system VD 0063 G AAA TCXX    Center of  Lifting point      gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 3:  Connections vacuum system VD 0063 G AAA TCXX      0870169324_VD 0025‐0100G_‐0006_IM_en  35 / 46  ...
  • Page 36     13 | Dimensions sheets  Dimensions vacuum system VD 0063 G AAA TCAX (mobile)    Fig. 21:  Dimensions vacuum system VD 0063 G AAA TCAX (mobile)    Center of        gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 4:  Connections vacuum system VD 0063 G AAA TCAX (mobile)    0870169324_VD 0025‐0100G_‐0006_IM_en 36 / 46   ...
  • Page 37       13 | Dimensions sheets  Dimensions vacuum system VD 0100 G AAA TDXX    Fig. 22:  Dimensions vacuum system VD 0100 G AAA TDXX    Center of  Lifting point      gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 5:  Connections vacuum system VD 0100 G AAA TDXX        0870169324_VD 0025‐0100G_‐0006_IM_en  37 / 46  ...
  • Page 38     13 | Dimensions sheets  Dimensions vacuum system VD 0100 G AAA TDAX (mobile)    Fig. 23:  Dimensions vacuum system VD 0100 G AAA TDAX (mobile)    Center of        gravity          Pos.  Designation  Connection  Standard  N1  Gas inlet  G 1 1/4“, female thread  ISO 228‐1  N2  Gas outlet  G 1 1/4“, female thread  ISO 228‐1  N3  Venting valve    Tab. 6:  Connections vacuum system VD 0100 G AAA TDAX (mobile)      0870169324_VD 0025‐0100G_‐0006_IM_en 38 / 46   ...
  • Page 39: Connection Values Of The Vacuum System Vd

          14 | Connection values of the vacuum system VD  14 Connection values of the vacuum system VD  14.1 Connection values  Vacuum system type  VD 0025 G  3 x 380 ‐ 400 V AC / PE  230 V AC  50 Hz  1.5 kW  3 A  16 A  VD 0063 G *  3 x 380 ‐ 400 V AC / PE  230 V AC  50 Hz  2.5 kW  5 A  16 A  VD 0100 G *  3 x 380 ‐ 400 V AC / PE  230 V AC  50 Hz  3.0 kW  6 A  16 A    Tab. 7:  Connection values of the vacuum systems VD    ATTENTION!  Damage to the drives.  • * Before commissioning, the circuit breaker (‐101Q1) must be set to the rated current of  the motor, see circuit diagram in the switch and control cabinet.    0870169324_VD 0025‐0100G_‐0006_IM_en ...
  • Page 40: Technical Data

        15 | Technical data  15 Technical data    Unit  VD 0025 G  VD 0063 G  VD 0100 G  Nominal suction capacity (50 Hz)  m³/h  25  63  100  Final pressure  hPa (mbar) abs.  1,5  Rated output of motor (50 Hz)  kW  1.0  2.0  2.7  Connected load switch and control  kW  1.5  2.5  3.0  cabinet (50 Hz)  Sound pressure level according to  dB(A)  60  64  65  EN ISO 2151 (at 50 Hz)  Max. water vapor compatibility  hPa (mbar)  40  with gas ballast valve (at 50 Hz)  Water vapor capacity  kg/h ...
  • Page 41: Oil

    16 Oil    VM 100  VSA 100 *  VSC 100  VSB 100  100  ISO‐VG  100  100  100  Oil type  Mineral oil  Synthetic oil  Synthetic oil  Synthetic oil  Ambience temperature range [°C]  5 ... 35  5 ... 40  5 ... 40  5 ... 40  Part number 1 l packaging  0831 000 060  0831 163 968  0831 168 356  0831 168 351  Part number 5 l packaging  0831 000 059  0831 136 969  0831 168 357  0831 168 352  Check the name plate (NP, fig. 2) to see which oil the vacuum pump was filled with.  * for applications in the food industry (H1), corrosion protection, not suitable for      continuous operation  If necessary, an oil with a different viscosity can be used for unfavorable room temperatures.  Contact Busch Service for more information.  0870169324_VD 0025‐0100G_‐0006_IM_en  41 / 46  ...
  • Page 42         0870169324_VD 0025‐0100G_‐0006_IM_en 42 / 46   ...
  • Page 43: Eu Declaration Of Conformity

          17 | EU Declaration of conformity  17 EU Declaration of conformity  This EU declaration of conformity and the CE marking on the name plate apply to the machine in the scope of the delivery  by Busch. The manufacturer is solely responsible for issuing this declaration of conformity.  If the machine is integrated into a higher‐level machine system, the manufacturer of this system (if applicable, the com‐ pany operating the system) must issue a declaration of conformity for the higher‐level machine or system and affix the CE  marking.  Manufacturer  Dr. Ing. K Busch GmbH  Schauinslandstr. 1  DE‐79689 Maulburg  Declaration for vacuum system(s) of type: VD 0025 G, VD 0063 G, VD 0100 G  with serial number from 2021 … to 2022 …  was/were produced according to the following EU standards:   "Machinery directive" 2006/42/EG   "Directive on electromagnetic compatibility" 2014/30/EU   “RoHS2 Directive” 2011/65/EU, 2017/2102, Restriction of the use of certain hazardous substances in electrical and  electronic devices  and according to the corresponding standards.  Standard  Name of the standard  EN ISO 12100:2010  Safety of machinery – General principles for design, risk assessment and  risk reduction  EN ISO 13857:2019  Safety of machinery – Safety distances to prevent hazard zones being  reached by upper and lower limbs  EN 1012‐1:2010  Compressors and vacuum pumps – Safety requirements – Parts 1 and 2  EN 1012‐2:1996+A1:2009  EN ISO 2151:2008  Acoustics – Noise test code for compressors and vacuum pumps –  Engineering method (grade 2) ...
  • Page 44               0870169324_VD 0025‐0100G_‐0006_IM_en 44 / 46   ...
  • Page 45     Notes...
  • Page 46 United Arab Emirates Chile service_sales@busch.co.il busch@busch.pt sales@busch.ae info@busch.cl Italy Romania United Kingdom China info@busch.it office@buschromania.ro sales@busch.co.uk info@busch-china.com Japan Russia Colombia info@busch.co.jp info@busch.ru info@buschusa.com info@buschvacuum.co Korea Singapore Czech Republic busch@busch.co.kr sales@busch.com.sg   info@buschvacuum.cz   www.buschvacuum.com 0870S169324/-0006_en / © Dr.-Ing. K. Busch GmbH...

Table of Contents