Updating Pipette Settings

Here you can set the following configurations on the connected PIPETMAN M pipette:
Set the Beep to On or Off (switches the tone of the pipette to On or Off). Beep On
and Off only affects the built-in pipette protocols, as well as Aspirate , Dispense,
Mix , Purge tasks (but NOT beep task!).
Set the Lock pipette On or Off (Lock pipette On disables access to Configuration,
Speed and Mode on PIPETMAN M pipette).
Set Maximal volume limit: this sets the maximal allowed volume on the particular
pipette; it cannot be higher than pipette's specified maximal volume. Simply enter
the volume (in µL) in the field.
Update Firmware.
Figure 26
Update pipette settings.
To modify these configurations, follow these steps:
1.
Click Update pipette settings in the main menu.
2.
If a pipette is already correctly connected, proceed to step 4.
3.
Connect pipette to the PC via USB cable and click Next. Should the pipette not be
recognized immediately, disconnect the USB cable, reconnect it and click Try again
button.
44
PIPETMAN M SOFTWARE
|
PIPETMAN® M