5.5

Configuration of the switch-off threshold

A threshold at which the fan of the marVac
consumption. This threshold can be configured using the L parameter
To do this, open the
setup menu item:
Initial state
Setup menu for line current detector
Submenu for activation/deactivation of
the line current detection function
(here: line current detection function
activated)
The switch-off threshold is increased or decreased using the Up and Down buttons.
The switch-on threshold is automatically calculated and cannot be configured.
Example
The connected surgical device has been calibrated with a quiescent current of 100 mA.
The L parameter has been configured at 20%.
The switch-off threshold, at which the fan absolutely must be deactivated, is 120 mA.
Revision 04
®
is switched off is calculated based on the measured power
setup menu ( see section 5.1, "Setup menu," page 42) and select the 4th
Button
3 x
Press the basic
suction button 3x
in succession
setup menu.
Objective
Setup menu for line current detector
Submenu for configuration of the switch-
off threshold (lower threshold) in percent,
relative to the measured quiescent
current consumption. The displayed
numerical value can be changed by the
Up/Down buttons within the permissible
limits (here: 15%)