Configuration Of The Switch-On Delay Time - KLS Martin group marVac Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 32
If the connected surgical device has been replaced or there have been significant changes in the
configuration of the connected surgical device, the calibration process must be repeated to avoid
malfunction.
The calibration process is started by pressing and holding the Up and Down buttons simultaneously for at
least 1 second:
Initial state
Setup menu for line current detector
Submenu for line current measurement (here:
calibration required)
Different parameters are possible instead of the
display. See section 5.1, p. 42.
A counter appears in the lower display segment counting down the seconds. The counter may reset itself
up to 3 times due to power fluctuations in the connected surgical device. Ideally, the calibration process
takes 15 seconds, in the worst case 60 seconds.
If the counter has expired, either the measured power consumption or an error code is displayed. See
section 5.1, "Setup menu," page 42.
The connected electrosurgery or laser surgery device must not be activated or operated during the
calibration process!
5.4

Configuration of the switch-on delay time

Fluctuations in the power consumption during the calibration process of the connected electrosurgery or
laser surgery device may cause the fan of the smoke evacuator marVac
can be prevented by configuration of a switch-on delay time of the current detector.
48
marVac® / maXium® smart Vac iDetect external line current detector
Button
Objective
Up + Down
at the same time
Setup menu for line current detector
for at least 1 s
Submenu for current measurement
(here: calibration in progress)
Instructions for Use
®
to be inadvertently activated. This
Revision 04

Advertisement

Table of Contents
loading

This manual is also suitable for:

Maxium smart vac

Table of Contents