Table 2-1. Exposure Signals Between PXCM and Generator
Signal
PREP_DR_DET_
PREP_DR_DET_
GEN_RDY_FOR_DR_EXP_
GEN_RDY_FOR_DR_EXP_
DR_DET_RDY_FOR_EXP_
DR_DET_RDY_FOR_EXP_
February 5, 2004
2-6
Direction
Generator to PXCM
Generator to PXCM
Generator to PXCM
Generator to PXCM
PXCM to Generator
PXCM to Generator
Description
The Generator has started the preparations to perform an
exposure.
Activation:
The Generator activates the signal when the exposure
preparations have started.
Deactivation:
The Generator deactivates the signal when:
The exposure buttons are released.
A correctly terminated exposure has been proceeded.
Deactivation:
The Generator deactivates the signal when:
An error interrupts the exposure.
For timing requirements, see t_prep_end_dly.
The Generator is ready to expose.
Activation:
When the anode speed is reached, the signal is activated.
Deactivation:
The Generator deactivates the signal when:
The exposure buttons are released.
A correctly terminated exposure has been proceeded.
Deactivation:
The Generator deactivates the signal when:
An error interrupts the exposure.
For timing requirements, see
t_gen_rdy_disable_to_ray_off.
Note: For an AEC pre-exposure, the signal will stay active
in the time interval between the AEC pre-exposure and the
AEC main exposure.
The Detector is ready for an exposure.
Activation:
When the Detector is ready to receive X-rays, the signal is
activated.
Deactivation:
When the Detector exposure window ends, the signal is
deactivated.