Hide individual waveforms in Wave View

Occasionally one waveform may obscure detail in another waveform. You can now selectively hide any
wave by clicking on the wave name (e.g. rate or kV) as listed at the top of the Wave graph. Clicking again
will un-hide the wave.
Wave view with difficult to see details
kV Off

Waveform filters

Measurements made with low-level signals (e.g. low mA, fluoro, scatter & small ion chambers) may display
noisy waveforms that obscure important details. Accu-Gold Wave View provides a dropdown selection to
apply a "Low-pass Bandwidth Filter" to the waveform. The CT exposure you see below was made with a
Radcal 0.15 cc thimble chamber placed in the central axial location of the Boone ICRU CT phantom. The
scan started and ended with the beam beyond the end of the phantom.
Dose Rate Off