Dd - Solid State Dose Sensors - Radcal ACCU-GOLD2 User Manual

For use with all accu-gold digitizers and accu-gold touch pro
Hide thumbs Also See for ACCU-GOLD2:
Table of Contents

Advertisement

DDX6-W (DISCONTINUED)
The DDX6-W is a solid-state dose sensor for diagnostic range measurements. It is a small solid-state
sensor for single exposures and fluoroscopy in the diagnostic range. It measures dose, dose-rate, and
time and is passively corrected for beam quality. It may be used simultaneously with an AGMS-D multi-
sensor in entrance / exit exposure measurements.
Sensor
Min Rate
Max Rate
Min Dose
Max Dose
Accuracy
Energy Dependence
Filtration Dependence
DDX6-W
2 R/s
23 R/s
10 R
64 kR
±5% using X-rays @ 80kVp with 2.5 mm Al total filtration (IEC
61267 RQR-6)
+5% from 50 kVp to 120 kVp at 2.5 mm Al
+5% to -10% from 2.5 mm Al to 23 mm Al
20 nGy/s
205 mGy/s
100 nGy
559 Gy