Sony LMD-2735MD Manual page 8

Hide thumbs Also See for LMD-2735MD:
Table of Contents

Advertisement

Guidance and manufacturer's dEiclaration - electromagnetic immunity
The LMD-27 35MD/ 2435MD is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. Portable RF communications equipment should be used no closer than 3 0 cm (1 2 inches) to any part of
the LMD-2735MD/2435MD. Otherwise, degradation of the performance of this equipment could result.
Immunity lest
380-390MHz
43 0-470MHz
70 4 -78 7 MHz
800-960MHz
Proximity fields
from RF wireless
communications
equipment
IEC 61 000 - 4-3
1,700-1,990 MHz
2,4 00 - 2,570 MHz
5,100-5,800 MHz
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a For some services, only the uplink frequencies are inclluded.
Band
Service"
a
TETRA4 00
GMRS 4 60
FRS 4 60
LTE Band 13„ 17
GSM 800 /9001
TETRA 800
iDE N 82 0
CDMA8 5 0
LTEBand 5
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3, 4,
25
UMTS
Bluetooth
WLAN
802. 11 b/g/n
RFID 245 0
LTEBand 7
WLAN 802. 11 a/n
IEC 60601
Modulation
test level
Pulse modulation
27V/m
18Hz
FM
kHz deviation
28 V/m
±5
1 kHz sine
Pulse modulation
9V/m
217 Hz
Pulse modulation
28 V/m
18 Hz
Pulse modulation
28 V/m
217 Hz
Pulse modulation
28 V/m
217 Hz
Pulse modulation
9V/m
217 Hz
Compliance level
27V/m
28V/m
9V/m
28V/m
28V/m
28V/m
9V/m

Advertisement