Troubleshooting - Ultrawave ARGON Operator's Instruction Manual

Ultrasonic cleaning system
Table of Contents

Advertisement

year or more for the full term of the extended warranty, as designated by Ultrawave for
each individual system/equipment.
Warranties will not apply if any fault has been caused by failure of the Customer to carry
out routine maintenance, in addition to any PPMS carried out by Ultrawave as part of
the extended warranty, or the Customer fails to ensure the required services are
maintained.
Ultrawave will accept no liability for any loss of production or business interruption
resulting from any system/equipment failure whether or not any such failure is covered
by any Ultrawave warranty.
Warranties are not transferrable to other users in the event of the system/equipment
being sold or otherwise transferred to third party users. Warranties will become void in
the event of the system/equipment being moved in any way from any location where
the system/equipment was originally commissioned by Ultrawave unless Ultrawave are
notified in writing prior to any such move.

Troubleshooting

Ultrawave have a dedicated After Sales team who are able to resolve any problems that
occur with your Argon Ultrasonic Cleaning System. However, on many occasions, it is
possible that the problem can be rectified by the operator.
The unit fails to turn on
(no display is shown)
The screen displays:
"Level Low"
The screen displays:
"Power Fail"
The screen displays:
"Under temp"
The screen displays:
"Over temp"
The screen displays:
"Sensor Fail"
If any of these problems persist, the Ultrawave After Sales Department can be
contacted on +44 (0) 29 2083 7337 or service@ultrawave.co.uk.
There are no user serviceable parts inside the Argon Series ultrasonic cleaning
Check that the system is plugged in and that mains
electricity is present.
Fill the tank with liquid so that it is above the float switch
inside the tank.
If the screen still displays "LEVEL LOW", there is an
internal fault and the ultrasonic cleaning system will need
to be serviced.
The system had lost power during a cycle, this will be
indicated at next power-up. This is just a notification to
the user that the last cycle may not have completed.
Allow the system to heat before starting another cycle
Allow the system to cool, or remove some of the hot
liquid and replace with cold liquid.
Switch the system off and then back on again. If the
screen still displays "Sensor Fail", there is an internal fault
and the system will need to be serviced.