Changing/Inspecting The Filters - Ultrawave ARGON Operator's Instruction Manual

Ultrasonic cleaning system
Table of Contents

Advertisement

Changing/Inspecting the filters

There are two valves used as isolation devices when changing the filters on the dual
filter system shown here in the normal open position:
Note: Do not shut off either valve when the Argon Series is in operation as this will
damage the pump.
The tank process and weir areas do not need to be emptied to change the filters.
Step 1 – Ensure that the drip tray is in place below the filter housings:
Step 2 – Turn both isolation valves 90 degrees to the "off / "closed" position to isolate
the tank fluid:
Step 3 – Open the filter housing bleed valve on the bottom of one of the housings:
Step 4 – Dispose of all fluid in the filter housing.
Step 5 – Repeat the same process with the second filter housing.