5. Data Output

5.2.3 To Transfer Raw Data (measurement data per second)

When transferring data, make sure to display the monitor screen.
While transferring data, do not use other functions.
[Output
Example]
Command
To
number of data
*
D
to be retrieved.
To turn probe
C
power OFF
To turn probe
S
power ON
* Command C and S can be used when you want to collect data every one hour or so.
By turning the probe off when it is not being used, unnecessary battery drain can be used.
* When command or number of data is entered incorrectly, the error codes listed below will be returned.
(Example: When the number of data exceeds the maximum, E2 will be returned in stead of AD.)
E1 ----- Command Error
E2 ----- Data Setting Error
www.lazodecontrol.com
Air Velocity
Air Temp.
23.6;
126;
Command Explanation
Function
set
the
After command is received, AD will be returned.
Then, the configured number of data (*) will be output every one second.
The maximum number of data to be output is 1,000 data. To retrieve more than 1,000 data,
send the command again.
After command is received, AC will be returned.
At the same time, the air velocity sensor will be turned off.
After command is received, the air velocity sensor will be turned on again.
30 seconds later AS will be returned and the instrument is ready to retrieve data.
After AS command is received, transfer data by sending D* command.
When SLOW is indicated, moving average
deviations will be transferred.
Please switch to monitoring screen.
Output Format
: 19 characters
dummy
: (unit: m/s,
*** CRLF
: Air Velocity, Air Temp ***
Explanation
B3 ----- Battery Error (battery drain)