3. Measurement Mode

3.5 Memory Capacity

Measurement data is stored per page. On top of the each page the
measurement condition is stored. Therefore, the more pages you
save, the less data you can store.
Below describes how to figure out how many pages of data can be
stored when you know how many data you want to take.
Page 1
(1) Measurement Starting Date (08. 7. 1 10:40)
(2) Measurement Item (air velocity, air temp VT)
(3) Measurement
(Average: AVE, Interval: INT, Air Flow: FLW)
(4) Sampling Time (S-TIME(S))
(5) No. of Data (DATA (N))
(6) No. of Measurement Points (POINTS)
(7) Measurement Interval INT (min) or Duct Area AREA (m²)
(8) Air Velocity Data and Air Temperature Data
Internal Memory
P = 4770 / (24 + 4 × N)
[Example] When N is 10, P is 74 according to the above formula.
www.lazodecontrol.com
Page 2
Page N
Number of Bytes Used per Page
How to Figure Out How Many Pages Can Be Saved
Memory Capacity 4770 bytes
Data
Measurement
Condition
Memory
Capacity
This means that when the number of data is 10, you can perform a measurement for 74 times
(pages).
★ Measurement data is stored per page.
★ Make sure to check in which page the
★ A measurement data consists of air
Measurement
Condition
Measurement
Data
12 bytes
2 bytes
2 bytes
2 bytes
2 bytes
2 bytes
2 bytes
4 bytes × No. of data
1 page
Average Value Measurement:
Intermittent Operation:
Air Flow Measurement:
+51 960 810 089
data is stored.
velocity data and air temperature
data.
Measurement Condition
Total: 24 bytes
(24+4×N) bytes
Data Volume (DATA (N))
Number of Points (POINTS)
Number of Points (POINTS)
ventas@lazodecontrol.com