2. Basic Operation
2.2 How to Make the Reading More Readable
This function is useful when you want to obtain averaged readings as readings were volatile.
* This function is not available in Average Value Measurement Mode, Intermittent Operation Measurement Mode
and Air Flow Measurement Mode.
Mode
0
FAST
0
SLOW 1
0
SLOW 2
www.lazodecontrol.com
Press FAST/SLOW key.
★ Every time you press FAST/SLOW key, the display at the
upper right corner changes in sequence.
FAST
SLOW 1
SLOW 2
How to Take in Measurement Data
5
10
15
average for 1 sec
5
10
15
20sec
average for 5 sec
5
10
15
20sec
FAST:
Instantaneous value every second
SLOW1: Moving average deviation for 5 sec
SLOW2: Moving average deviation for 10 sec
20sec
(measuring time)
Data is taken 4 times for one second. The
average value of the 4 data is displayed as
an instantaneous value every second.
(measuring time)
The average value for 5 seconds is
displayed every second.
Data shifts by 1 second.
(measuring time)
The average value for 10 seconds is
displayed every second.
average for 10 sec
Data shifts by 1 second.