Troubleshooting - Flaem Pro OxyLast Med Easy P0613EM User Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

Directions:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable water and disin-
fectant, according to the proportions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the
parts. Leave the parts immersed for the amount of time indicated on the packaging of the disinfectant
associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- After having disinfected the accessories, shake them vigorously and place them on a paper towel, or dry
them with a jet of hot air (for example, from a hair dryer).
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
STERILISATION
Accessories (6a/6b/6c) can be sterilised.
The tubes, filter, connectors and all fittings CANNOT be sterilised as the polymers they are made of cannot
withstand high temperatures.
The sterilisation procedure described in this paragraph is effective on the parts that undergo this treat-
ment only if every step is followed and only if the parts to be treated have previously been sanitised. The
procedure is validated in its conformity to ISO 17665-1.
Device: Steam steriliser with fractionated vacuum and overpressure in accordance with EN 13060.
Directions: Wrap every single part to be treated with a sterile barrier system or packaging in accordance
with standard EN 11607. Place the packed components in the steam steriliser. Run the sterilisation cycle
according to the operating instructions of the device by selecting a temperature of 134°C and a time of 10
minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile barrier system or pack-
aging.
TROUBLE-SHOOTING
Problem
LED 1.1 POWER does not switch
on
The LED 1.1 POWER is on but the
device does not work
The LED ALARM 1.7 is on
Cause
The power supply cable is not
correctly attached.
Fuses trip
Circuit breaker tripped
Overpressure protection
tripped
17
Solution
Insert the power cable.
Take it to an authorised centre.
Allow the medical device to cool
for at least 40 minutes.
Try and switch it back on and if
it still does not work take it to an
authorised centre.
Take it to an authorised centre