TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
The device does not work
The device does not
nebulize or
nebulizes insufficiently.
The apparatus is noisier
than usual
If after verifying the above mentioned conditions the device should not operate
properly, contact your trusted retailer or the nearest authorized FLAEM service center.
CAUSE
The power cable has not been
correctly inserted into the
socket of the device or the
power socket
The medication has not been
inserted in the nebulizer
The nebulizer has not been
properly fitted.
The nebulizer nozzle is
clogged
The air pipe is not correctly
connected to the device
The air pipe is bent, damaged
or twisted
The air filter is dirty
The Filter is not correctly
inserted in its housing
14
SOLUTION
Correctly insert the power cable
in sockets.
Pour the right amount of
medication
in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on
the cover.
Disassemble the nebulizer,
remove the nozzle and perform
cleaning operations Failure to
remove medication deposits
from the nebulizer affects its
efficiency and operation.
Strictly comply with the
instructions contained in the
"CLEANING, SANITISATION,
DISINFECTION" chapter
Ensure the device's accessories
are properly connected to the air
intake (see connection diagram
on the cover).
Unwind the pipe and ensure it is
not crushed or punctured. If
necessary replace it.
Replace the filter
Fully and properly insert the
Filter in its housing
EL
Th
co
ins
an
po
tel
wi
ww
if o
ma