9

Supplement

9.1

Technical data

General data
Series
Weight
Housing material
Socket material
Voltage supply
Operating voltage
Max. power consumption
Sensor input
Quantity
Data transmission
Hysteresis
Switching threshold
Current limitation
Sensor operating voltage
Detection line break
Detection shortcircuit
Configuration, connection cable
Resistance per conductor
Relay output
Number, function
Switching delay
Mode
Contact
Contact material
Switching voltage
Switching current
Breaking capacity
Transistor output
Number, function
Galvanic separation
Maximum values
Ʋ U
B
Ʋ I
B
Transistor voltage loss (U
VEGATOR 636 Ex • Signal conditioning instrument
Module unit with plug-in socket for mounting on carrier
rail 35 x 7.5 or 35 x 5 according to EN 50022
170 g (6 oz)
Noryl SE100, Lexan 920A
Noryl SE100, Noryl SE1 GFN3
20 ... 250 V AC, 50/60 Hz, 20 ... 72 V DC
3 W (3 ... 18 VA)
1
Analogue
100 µA
12 mA
24 mA (permanently short-circuit proof)
15 ... 18 V DC
< 2.3 mA
> 23.2 mA
2-wire
max. 35 Ω
1 x switching relay (spdt)
0.2 ... 20 s, directional switching
A/B switch (A - max. detection or overfill protection, B -
min. detection or dry run protection)
1 x spdt
AgNi 0.15 hard gold-plated
≥ 10 mV DC, ≤ 253 V AC/DC
≥ 10 µA DC, ≤ 3 A AC, 1 A DC
≤ 500 VA, ≤ 54 W DC
1 output, synchronously switching with the relay
Floating
36 V DC
≤ 60 mA, short-circuit proof
)
approx. 1.5 V at I
CE
60 mA
B
9 Supplement