Order Of Cell Components; Control Of Cell Purity And Absorbing Mobile Phase - Ecom ECD2800 User Manual

Uv-vis detector
Table of Contents

Advertisement

14.1. Order of Cell Components

1. Screw M2x4 07608, 2. Cell lid ADA53020, 3. Cell lid seal ADA53030, 4. Cell window 4mm 78000110,
1. Screw M2x10 17878, 2. Microcell lid ADA61020, 3. Microcell lid seal ADA61030,
4. Cell window 4mm 78000110 , 5. Cell body 2mm PEEK, 6. Microcell nut ADA61040

14.2. Control of Cell Purity and Absorbing Mobile Phase

Control of cell purity and absorbing mobile phase is performed at the wavelength used for
measuring or at the wavelengths used in the program.
Caution:
Measure on a stabilized unit and measuring cell without bubbles.
A common problem are also air bubbles stuck inside the cell in the area of light transmission.
They cause worse measuring results. The bubbles may be caused by gas released in the liquid.
Stuck bubbles may be seen with naked eye after removing cell from the unit and its observation
through optical part against light (e.g. window, etc.). Appropriate way how to remove them is to
rinse the cell with HPLC acetone or dry it with nitrogen gas.
P
ARTS OF ANALYTICAL CELLS
5. Cell body – PEEK.
P
ARTS OF ANALYTICAL CELL
37 / 47
ECD2800 / ECD2600
AC05, AC10