Info; Device; Boards; Ethernet - Ecom ECD2800 User Manual

Uv-vis detector
Table of Contents

Advertisement

Negative dynamic range setting does not influence output range of analogue output. If there
is a value −0.8 AU (neg. dynamic range = HIGHER) on the display, on analogue output will be
minimal value for selected range.
6.4.4.1.
Range
Here can be set output voltage range in four levels (1, 2, 3 and 4 V)
6.4.4.2.
Offset
Offset of analogue output can be set in range 0 – 1 V (default value is 0 mV)
6.4.4.3.
Conversion Ratio
Conversion ratio of output voltage to AU can be set in six ranges 0.2/0.5/1/2/5/10 V = 1 AU
(default value is 1 V/AU).

6.5. Info

6.5.1. Device

Displays unit NAME, its S/N - serial number and TIME - total operating time.

6.5.2. Boards

Here can be selected board we need information about
6.5.2.1.
Main Board
BN – board name, HW – board version, FW – firmware version, UN – board identification
number.
6.5.2.2.
Display Board
BN – board name, HW – board version, FW – firmware version, UN – board identification
number.
6.5.2.3.
Power Board
BN – board name, HW – board version, FW – firmware version.
6.5.2.4.
XPORT Module
Ethernet module..
BN – board name, HW – board version, FW – firmware version, UN – board identification
number.

6.5.3. Ethernet

Actual setting of Ethernet. IP – IP address, MASK – subnet mask, GATE – default gateway,
MAC – unique hardware address.

6.5.4. User text

Default is empty. The user may input supplementary information about the detector when
using software. E.g. location in the laboratory. USER TEXT is then a part of the protocol.