If an error is reported during calibration, original values will stay recorded, which may cause
that maximal value may differ from 100 % value.
Light intensity is the quantity of light incident on reference (REF, prior to cell) or measuring
(SAM, after the cell) sensor. Value is presented in percentage and 100 % is set for maximal
intensity of new lamp. This value is spectrally dependent (see following graphs) and during light
aging it decreases. Utmost decrease is in the area of low UV. This value is used as benchmark
data about light intensity on the sensors or for cell purity diagnostics and is displayed on the
absorbance measuring screen. For example on wavelength 254 nm will be displayed: R52%
S53%. If S value is significantly lower, it denotes impurities in the cell or absorbing mobile phase.
120
100
80
60
40
20
0
190
D2 = deuterium lamp, W = tungsten lamp, second order filters not inserted
Confirm Replacement
When entering this part of menu and confirming new lamp, operating hours are set to zero.
Data about old lamp are archived in the unit memory.
6.7.1.2.
Tungsten Lamp (W)
G
L
RAPH OF
IGHT
290
390
490
Wavelength [nm]
I
N
L
NTENSITY WITH
EW
590
690
Disconnect the unit from power
source!!!
Unscrew the lamp cover with
screwdriver Torx T10
30 / 47
ECD2800 / ECD2600
AMP
D2 REF
D2 SAM
D2+W REF
D2+W SAM