Table 4-7. Calibration Settings Dialog Box
Item
TES Name
1
Test_Mod_Frequency_H
B
2
TLS_PowerLevel_HB
3
VNA_AverageFactor_HB
4
VNA_IFBandwidth_HB
5
Cal_Verify_Limit
6
TLS_Wavelength
7
TLS_PowerLevel
8
VNA_AverageFactor
9
Refractive_Index
10
Cal_Len_MaxLength
Parameter Location
Group: New_ Calibration
Test: Cal_Gain_Phase
Group: Verify_Calibration
Test: Verify_Gain_Phase
Group: New_ Calibration
Test: Cal_Gain_Phase
Group: Verify_Calibration
Test: Verify_Gain_Phase
Group: New_ Calibration
Test: Cal_Gain_Phase
Group: Verify_Calibration
Test: Verify_Gain_Phase
Group: New_ Calibration
Test: Cal_Gain_Phase
Group: Verify_Calibration
Test: Verify_Gain_Phase
Test Plan:
Standard_Test_Plan
Group: New_ Calibration
Test: Cal_Length
Group: New_ Calibration
Test: Cal_Length
Group: New_ Calibration
Test: Cal_Length
Group: New_ Calibration
Test: Cal_Length
Test Plan:
Standard_Test_Plan
Type
Comments
Single
Single
Integer
Integer
This parameter is "Sensitivity" on the GUI,
but the tests need an IF bandwidth. The IF
bandwidth number is stored in the database.
Normal = 300. High = 100. Highest = 30.
Single
This number is stored in seconds in the
database.
Single
Single
Integer
Double
Single
Remote Control