Download Print this page
Agilent Technologies IDP3GBKIT Quick Reference Manual

Agilent Technologies IDP3GBKIT Quick Reference Manual

Advertisement

Quick Links

Quick Reference Guide
TPS-compact Gas Ballast
The TPS-compact packaging includes a standard gas ballast kit (P/N IDP3GBKIT) that can be used to prevent water and
other condensates from accumulating within the foreline pump. The standard configuration has a seal screw installed in the
1/4-20 thread in the top of the exhaust manifold. This configuration can be used for dry applications.
When appreciable amounts of water, solvents, etc., are being pumped, remove the seal screw and replace it with the provided
adapter and 1/8" NPT sintered filter plug.
Before install gas ballast adapter remove the power supply cable.
WARNING
1. Disconnect the TPS-compact from electrical power.
2. Unlock the scroll pump support by unscrewing the screws.
3. Lift the TPS-compact in order to access the Scroll pump site.

Advertisement

loading

Summary of Contents for Agilent Technologies IDP3GBKIT

  • Page 1 TPS-compact Gas Ballast The TPS-compact packaging includes a standard gas ballast kit (P/N IDP3GBKIT) that can be used to prevent water and other condensates from accumulating within the foreline pump. The standard configuration has a seal screw installed in the 1/4-20 thread in the top of the exhaust manifold.
  • Page 2 +00 800 234 234 00 400 8206778 (mobile) vpl-customercare@agilent.com vpt-customercare@agilent.com 800 8206778 (landline) vpd-support-china@agilent.com Visit the product web page: www.agilent.com/chem/tpscompact This information is subject to change without notice. © Agilent Technologies, Inc. 2022 Printed in Italy, January 14th, 2022 X3582-90000-9-PUI rev. A.00...