3. TECHNICAL DESCRIPTION

3.1AM5000 AM5000/3 - AM5000/5 anaesthesia module
AM5000 - AM5000/3 – AM5000/5 regulates the supply and concentration of the gas mixture (Air, O
and also supplies it to the anaesthetic gas vaporizer.
It also allows the selection of the gas mixture to be supplied (Air-O
oxygen) of the supplied gas mixture for emergencies. The AM5000 - AM5000/3 – AM5000/5 is also equipped
with the MIX-LIFE device that guarantees a minimum of 25% oxygen in all open-tap conditions.
By using the three gauges located on the front panel, the medical gases supply pressure coming from the
central gas installation can be continuously controlled (precision ± 10%).
The flowmeters permit the measurement of the flow the of the corresponding gases with a precision of ±
10% of the displayed value or ± 1% of the bottom of the scale, by choosing the highest of the two values.
AM5000/3
L
I
I
OXYGEN supply gauge.
L
NITROUS OXIDE supply gauge.
M
AIR supply gauge.
N
Selector switch for AIR / NITROUS OXIDE. This safety device:
avoids the contemporaneous supply of air and nitrous oxide
only the selected gas will be available on the flowmeters.
O
OXYGEN BY-PASS key. By pressing this key pure oxygen is released into the anaesthesia circuit with a
flow of approximately 60 l/min.
P
OXYGEN flow regulator for fresh gases. It opens counter-clockwise.
Q
NITROUS OXIDE flow regulator for fresh gases. It opens counter-clockwise.
P.N.: DU5506130
‫.ﻣراﺟﻌﮫ ﻓرﻣﺎﯾﯾد‬WWW.ELM-FAN.IR‫ﺑرای داﻧﻠود ﻣوارد ﺑﯾﺷﺗر ﺑﮫ ﺳﺎﯾت‬
AM5000 – AM5000/3 – AM5000/5
AM5000
I
L
S T
U
M
N
O
P
Q
V
Page 6 of 34
To download more items, go to ELM-FAN.IR
, or N
2
S T
O
P
Q
V
AM5000/5
I
L
M
N
O
R
Front view
Service Manual
O-O
) and the enrichment (with
2
2
S
U
W
T
X
R
P
Q
V
, N
O),
2
2
Rev.: 3.0

Advertisement

loading

This manual is also suitable for:

Am5000/3Am5000/5