SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
Service Indicator Light Diagnostics
1
E 01
User Fault
2
E02
Battery Fault
3
E03
Left Motor Fault
4
E04
Right Motor Fault
Pronto® M71™
All manuals and user guides at all-guides.com
00
Stall Timeout or user error.
00 Recharge batteries or replace.
00
Left Motor Short Circuit
01
Left Motor Open Circuit
02
Left Motor Connection Fault
B-
03
Motor Terminal Connected
to B+
04
Left Motor Voltage Fault
05
Left Motor Bridge Fault
06
Too Many Hardware Current
Limit Events
07
Current Offset Out of Range
08 Hardware Current Limit Fault
00
Right Motor Short Circuit
01
Right Motor Open Circuit
02 Right Motor Connection Fault
B-
03
Motor Terminal Connected
to B+
04
Right Motor Voltage Fault
05
Right Motor Bridge Fault
06
Too Many Hardware Current
Limit Events
07
Current Offset Out of Range
08 Hardware Current Limit Fault
20
Release joystick to neutral and try again.
Check the batteries and cable. Try
charging the batteries. Batteries may
require replacing.
Check the left motor, connections and
motor cable.
Check the right motor, connections and
motor cable.