Gait up GO Instructions For Use Manual page 27

Hide thumbs Also See for GO:
Table of Contents

Advertisement

level of the Physilog6 sensor device on the Gait Up GO Mobile Application screen before starting any
test (see section 4.5.5).
● Plug one end of the USB cable (USB-C end) into the Physilog6 sensor device USB-C port.
● Plug the USB-A end of the UCB cable into a computer or a mobile phone charger with
specifications compatible to the ones listed under section 9 of the present document,
● During charging, the LED on the Physilog6 sensor device emits a blinking white light, even
when the sensor device is not connected to the Mobile Application.
● When the charging is complete, the LED on the Physilog6 sensor device will shine green and
the battery level shown in the Gait Up GO Mobile Application will appear as full.
It is important to charge the Physilog6 sensor device for at least 4 hours before an assessment session
is started.
Note: It is recommended to charge the Physilog6 sensor device at the end of each session.
Note: Use only the USB cable provided by the manufacturer.
Note: Use only mobile phone chargers with specifications compatible to the ones provided by the
manufacturer in section 9 of the present document.
Note: Please, in order to keep the performance of the device, recharge the battery of the device every 3
months if it is stored and not used
4.3.4 Checking system components and their functionality
● Visually inspect all the components of Gait Up GO medical device (Mobile phone/tablet,
Physilog6 sensor device, clip accessories and USB cable) for signs of damage, especially in
correspondence of connectors, primary packaging (casing) where the Physilog6 sensor device
and the clip accessories are contained.
Note: In case of lack of integrity of one or more components of the Gait Up GO medical device,
please immediately contact the customer support.
● Clean the Physilog6 sensor device and the clip accessories before using the device. See section
5. Cleaning, Maintenance and Storage of the present document.
● Make sure that the Mobile Phone/Tablet is fully charged before starting an assessment session.
● Make sure that the Physilog6 sensor device is fully charged before starting an assessment
session.
● Do not operate the device while the Physilog6 sensor device is being charged.
● Open the Gait Up GO Mobile Application and connect it to the relevant Physilog 6 sensor
device. Check that there are no connection issues. Refer to the section 4.4.2 of the present
document for more information.
● Position the sensor device into the clip accessories and check that the Physilog6 sensor device
is firmly attached to the clip.
Note: In case of BLE connection issues, please contact the customer support.