ABB AV410 User Manual page 61

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

Relay outputs
Number of relays
Three supplied as standard,
configurable to one or both sensor inputs or status
Set point adjustment
Fully programmable as normal or failsafe, high/low or status
Hysteresis
Programmable 0 to 5% in 0.1% increments
Delay
Programmable 0 to 100 minutes in 1 minute intervals
Relay contacts
Single-pole changeover
Rating 5A 115/230V AC, 5A DC
Insulation
2kV RMS contacts to earth/ground
Power supply
Voltage requirements
85 to 260V AC 50/60 Hz
Optional 9 to 36V DC
Power consumption
<24VA
Insulation
Mains to earth (line to ground) 2kV RMS
Mechanical Data
Transmitter
IP66/NEMA4X
Dimensions 192mm (7.56 in.) high x 230mm (9.06 in.) wide x
94mm (3.7 in.) deep
Weight
1kg (2.2 lb)
Sensor
Low Range
Dimensions 327mm (12.87 in.) wide
x 410mm (16.14 in.) high x 162mm (6.38 in) deep
High Range
Dimensions 405mm (15.94 in.) wide
x 373mm (14.68 in.) high x 136mm (5.35 in) deep
Weight
6kg (13.2 lb)
Environmental Data
Operating temperature limits
0 to 50°C (32 to 122°F)
Storage temperature limits
–25 to 75°C (–13 to 167°F)
Operating humidity limits
Up to 95%RH non-condensing
EMC emissions and immunity
Meets requirements of:
EN61326 (for an industrial environment)
EN50081-2
EN50082-2
Safety
EN61010-1
Overvoltage Class II on inputs and outputs
Pollution Category 2
Cable entry types
Standard
5 or 7 x M20 cable glands
N. American
7 x knockouts suitable for
SPECIFICATION
1
/
in. Hubble gland
2
SS/AV4ORG Issue 3
59